CU-ERT Operator’s Manual_ver 1.10
40
iv.
Rhythm converted to non shockable Normal Sinus Rhythm.
v.
Refibrillation to Ventricular Fibrillation after 20 seconds.
vi.
2
nd
shock
vii.
CPR countdown (30:2, 15:2)
viii.
Rhythm converted to non shockable Normal Sinus Rhythm.
d.
The name of the Training Scenario will be shown on the LCD before the said scenario is displayed
on the LCD screen and while the scenario is running.
6.3.15 S9, Training Scenario 9
a.
This is initiated by pressing the S9 key on the Remote Controller
b.
Ventricular Fibrillation all throughout, no conversion to Normal Sinus Rhythm
c.
The name of the Training Scenario will be shown on the LCD before the said scenario is displayed
on the LCD screen and while the scenario is running.
6.3.16 S10, Training Scenario 10
a.
This is initiated by pressing the S10 key on the Remote Controller
b.
Ventricular Fibrillation with one shock required for conversion followed by a low battery condition.
c.
The following will be simulated:
i.
Shockable rhythm
ii.
1
st
shock
iii.
CPR countdown (30:2, 15:2)
iv.
Rhythm converted to non shockable Normal Sinus Rhythm.
v.
Low battery condition
d.
The name of the Training Scenario will be shown on the LCD before the said scenario is displayed
on the LCD screen and while the scenario is running.