FCS - Operation Manual
Rev. C | 2022.02
Enter the corresponding number for the analog output that you want to test and enter the
length of time you want to test the test to last. Enter the output value in mA that you want
tested. Press Start Test to begin the test.
To cancel the test, go back into the Test menu, open the Test Analog Output screen and enter
0 in the Minutes field and press Start Test.
6 BASIC SETTINGS AND CONFIGURATIONS
6.1 Factory Default Settings
The channel/gas operation type settings are configured at the factory according to the
specifications at time of order. The remaining settings are usually configured in the factory
according the default settings listed below. (There may be some differences in the default
settings depending on application.) If changes to the default settings are desired, the settings
can be changed in the field as indicated in the table below.
ITEM
DEFAULT SETTING
FIELD CONFIGURABLE SETTINGS
Door Mounted Buzzer
Alarm Level
Silence Interval
Enabled - Normal
High
300 seconds (5
minutes)
Enable, Disable - Normal, Failsafe
Low, Mid, High
0 to 9999 seconds
Strobe
ON Delay
Silence Interval
Enabled - Normal
10 seconds
300 seconds (5
minutes)
Enable, Disable - Failsafe, Normal
value is application dependant
0 to 9999 seconds
Data Logger
Enabled
5 minutes
Enable, Disable
1 to 1080 (18 hrs)
Relay 1
Relay 2
Relay 3
ON Delay (each relay)
OFF Delay (each relay)
Low - Failsafe
Mid - Failsafe
High - Failsafe
10 seconds
10 seconds
Low, Mid, High - Failsafe, Normal
Low, Mid, High - Failsafe, Normal
Low, Mid, HIgh - Failsafe, Normal
value is application dependant
value is application dependant