17
Section 1- Service Access and Calibration Procedure
1. Updating or Re-programming software
2. Front Service Panel Access
3. PCB Board/Control Panel Access
4. Scale/PC Board Recalibration Procedure
5. Tank Weight configuration
6. Scale Tare Procedure
7. Gauge Calibration Procedure
8. Pressure Transducer Recalibration
Note:
If there is any question on how to perform these procedures, please consult CPS service
center.
1.
Updating Program
a.
The unit can be updated by using USB stick or SD card.
b.
The SD card is simply inserted to the SD reader on the back of control panel and turn on unit. The unit will
beep three times at the start of installation, and then single beeps as it loads program. The unit will beep
three times at the end if down load. No password is needed. Do not remove SD card unit main screen “Do
Not Connect to Car” comes up.
c.
The USB is used from the update date screen. On the home screen push the gear icon on bottom of screen.
Two new icons will appear. Select the wrenches on right side of the screen. Then six new icons will appear.
Select the Up grade icon and install USB stick into USB Port. The screen will detect program and then push
green arrow and will prom you for password that came with update.