13
INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS
5.1.2 Pre-filter
Air entering the workstation is pre-filtered with an efficiency
of 83%.
WARNING
Only use original K-SYSTEMS VOC
filters.
5.1.3 Fan
The air is drawn into the fan in the top of the workstation
where it is pressurized. From the pressure plenum the air
passes through the HEPA main filter.
5.1.4 Main Filter
The filter efficiency of the HEPA main filter is 99.995% of particles 0.3 μm (D.O.P. test).
The air flows from the main filter through the work chamber in a vertical unidirectional flow of clean air.
Immediately before reaching the tabletop, the air separates and flows out through the back wall, and
through the work opening. The air returns to the suction opening of the workstation passing through the
surrounding space.