32
INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS
6.10.2 Start Set (St.St)
The St.St (Start Set) option is the
timer function
for the
HEAT
and the
FAN
. By using this option, the
workstation can be ready for use before procedures start in the morning.
For example
: If you want the heater to turn on at 8:00 AM but no fan, set the St.St (timer) to 8.00 and set
the HEAT to ON.
NOTE:
This function will only work correctly if the clock is set (ti.St).
Follow these steps to navigate in the Start Set (St.St) option.
ACTION
KEY
DISPLAY
Press and hold both the UP
and
DOWN keys for 3
seconds to enter the menu.
Press the DOWN key until the menu "-ti-" appears in the
display.
Press the SP key once to enter the time menu.
Press the DOWN key until the option "St.St" appears in
the display.
Press and hold the SP key to the change values.
While holding the SP key, press the UP key to change
the minutes and the DOWN key to change the hours.
When the time is set, release the SP key.
Press and hold both the UP
and
DOWN keys for 3
seconds to exit the menu.
St.St is connected with HEAT and FAN. It must be chosen when the timer starts. To turn the heater ON,
see the option HEAT/FAN (sub-sections 6.9.1 and 6.9.2).
The options are:
Heat (ON/OFF) and Fan (SLOW speed, FAST speed or OFF). These can be combined
with the timer in any way desired.
CAUTION
Always make sure when the timer function is being used, that the heated area is
clear of any objects that might be damaged by the heat or adversely affected by it
in any way. Caution should always be exercised when a heated area is turned on
without any supervision.
Summary of Contents for K-Systems L100 IVF Series
Page 1: ...K SYSTEMS IVF L100 Series Workstations Instruction Manual...
Page 76: ...76 INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS...
Page 77: ...77 INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS...
Page 78: ...78 INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS...
Page 79: ...79 INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS...
Page 80: ...80 INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS...