36
INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS
This Ultraviolet Light label should be posted outside laboratories containing
equipment capable of generating ultraviolet light above the NIOSH published
recommended exposure limit (REL) for occupational exposure. At 254 nm,
the wavelength for ultraviolet germicidal irradiation (UVGI), the REL is 0.006
joules per square centimeter (0.006 J/cm
2
). The permissible irradiance for an
8 hour workday exposure is <0.2 W/cm
2
or 6000 Wsec/cm
2
.
6.10.5 UV Light Off (u-oF)
This function is only activated if a UV light is installed in the workstation. This option sets a timer for the
UV light and the light will automatically turn OFF at "uv-oF".
Follow these steps to navigate in the UV Light off (uv-oF) option.
ACTION
KEY
DISPLAY
Press and hold both the UP
and
DOWN key for 3 seconds to
enter the menu.
Press the DOWN key until the menu "-ti-" appears in the
display.
Press the SP key once to enter time menu.
Press the DOWN key until the option "u-oF" appears in the
display.
To set the timer, press and hold the SP key.
Press the DOWN key to set the hour and the UP key to set
the minutes.
Press and hold both the UP
and
DOWN keys for 3 seconds
to exit the menu.
NOTE:
This function only works correctly if “ti.St “is set.
Summary of Contents for K-Systems L100 IVF Series
Page 1: ...K SYSTEMS IVF L100 Series Workstations Instruction Manual...
Page 76: ...76 INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS...
Page 77: ...77 INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS...
Page 78: ...78 INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS...
Page 79: ...79 INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS...
Page 80: ...80 INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS...