INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS
6.11.3. Calibrate Value Zone-2 (tn-2)
The tn-2 option is for the calibration of the temperature on Zone-2 (See Section 7: Different Heated
Surfaces).
If there is an offset between the value on the display and any measurements made with a high precision
external temperature sensor, this can be corrected. The new temperature reading will be kept as
the displayed value, with temperature control conducted on this basis (See Section 9: Temperature
Calibration).
Follow these steps to navigate in the Calibrate Value Zone-2 (tn-2) option.
ACTION
KEY
DISPLAY
Press and hold both the UP
and
DOWN key for 3 seconds
to enter the menu.
Press the DOWN key until the menu "StuP" appears in the
display.
Press the SP key once to enter the Setup menu.
Press the DOWN key until the option "tn-2" appears in the
display.
Press and hold the SP key to see the current value.
While holding the SP key, press either the UP key
or
the
DOWN key to select the calibration value.
When the required value has been selected, release the SP
key.
Press and hold both the UP
and
DOWN keys for 3 seconds
to exit the menu.
NOTE:
Always wait for the temperature to be stabilized before calibrating.
Summary of Contents for K-Systems L100 IVF Series
Page 1: ...K SYSTEMS IVF L100 Series Workstations Instruction Manual...
Page 76: ...76 INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS...
Page 77: ...77 INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS...
Page 78: ...78 INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS...
Page 79: ...79 INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS...
Page 80: ...80 INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS...