INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS
Dual 2xT845 (6xregulation)
Type: 2xT845
Size: 2 x 800 x 430 mm
Regulation: 6
The safe area is located 50 mm from the edge.
In this case there are three different temperature sensors on each side that may not be calibrated.
Each zone most be calibrated separately, to calibrate "Zone 1", measure only temperature in "Zone 1".
The red zone on the picture is "cross-heat" zone; do not measure the temperature in the cross-heat
zone.