A1000 INSTALLATION, OPERATION, & MAINTENANCE MANUAL
February 2016 | 256574-ENG
R01 6-1
6 TROUBLESHOOTING
Even if service is close by, a few troubleshooting steps can significantly reduce the time to
diagnose and correct a fault. Make careful notes of the fault symptoms and the chamber and
ambient conditions. This could help to determine the cause of the problem.
Chamber won’t start
1.
Check that the chamber is plugged into the wall outlet and that the main disconnect switch is ON.
2.
Check the circuit breaker at the building electrical panel.
3.
Ensure the program is set and running in the controller and the start/stop switch is ON.
4.
Check the temperature limit settings and ensure they are outside the program range.
Still won’t start
Contact service or Conviron.
Chamber won’t cool
1.
Check that the ambient temperature is below 30
C.
2.
Ensure that the door is firmly closed.
3.
Ensure that the fresh air intake is closed.
4.
Ensure that the temperature sensor is in the correct position.
Still won’t cool
Contact service or Conviron.
Chamber won’t heat
1.
Check that the ambient temperature is above 20
C.
2.
Ensure that the door is firmly closed.
3.
Ensure that the fresh air intake is closed.
4.
Ensure that the temperature sensor is in the correct position.
Chamber won’t make humidity
1.
Ensure that the door is firmly closed.
2.
Ensure that the fresh air intake is closed.
3.
Check the water supply to the rear of the chamber.
Still no humidify
Contact service or Conviron.
Conviron Technical Support is available to all users at no charge, to either assist with
troubleshooting or to order parts, for the life of the equipment.
Summary of Contents for ADAPTIS A1000
Page 1: ...A1000 INSTALLATION OPERATION MAINTENANCE MANUAL 256574 ENG R01 ...
Page 14: ...Precautions 1 4 February 2016 256574 ENG R01 ...
Page 46: ...Kit Installation 2 32 February 2016 256574 ENG R01 ...
Page 54: ...Chamber Start Up and Operation 5 4 February 2016 256574 ENG R01 ...
Page 63: ......