A1000 INSTALLATION, OPERATION, & MAINTENANCE MANUAL
February 2016 | 256574-ENG
R01 2-21
2.4.2 IN Kit Installation
Open the chamber door, remove the kit package, and inspect the chamber and kit parts. Contact
Conviron immediately is any damage is found, or if any parts are missing.
2.4.2.1 Installing the Back Wall Plenum Support
This support is very important since it closes off the bottom of the chamber plenum and directs the
airflow through the back wall plenum and into the chamber at equal pressure. If this support is
installed upside down, the back wall plenum will not fit. If the support is installed too loosely, the
airflow will not be evenly distributed through the back wall plenum.
To install the back wall plenum bottom support:
Install the support, Item A, onto the back wall of the chamber with four (4) screws, Item C (Figure
2-24) and tighten the screws to full resistance.
Figure 2-24
IN Back Wall Plenum Bottom Support Installed
The back wall plenum bottom support and the back wall plenum may be installed at
the factory.
Install the support with the large flange pointing upward as shown in Figure 2-24
C
A