Chamber Installation Requirements
3-2 February 2016 | 256574-ENG
R01
3.1.3 Water
Supply
Flow
0.26 gallons / hour (1 Liters / hour) purified water
Pressure
Max. – 60 PSI (4 bar), Min. – 5 PSI (0.3 bar)
pH
7.0 ± 0.5
Filtration
< 0.00008 inch (2 microns)
Resistance
0.5 to 5.0 Mega Ohms
Conductivity
2.0 to 0.2
μ
Siemens
3.1.4 Condensate
Drain
A ½” drain is provided underneath chamber, located in the center. The drain consists of ½” tubing
with p-trap and includes a drip pan for collection of condensate. This ½” drain may be extended to
a nearby floor drain, as required.
Install the drip pan if the connection to a floor drain is not required.
It is important to use a water source with the quality stated above, as failure to do so
will void the product warranty.