Communications
80
515 BR03 - 15 August 2021
Current and Logged Process Data
This block of registers is available for the retrieval of current or logged
process data with its matching time and date information.
Use the log timebase and log number to retrieve the logged information
from the appropriate register. If a particular log number does not exist, or
the instrument does not have the optional real-time clock, the time and date
stamp and associated variables are set to zero.
Register
Name
Comments
Read Only or
Read/Write
Type
1
Net Mass
Process Variables
By default totals are the Accumulated values. If
current Non-accumulated (resettable) totals
are required, set register 37 to 06.
R
P
3
Net Flowrate
R
P
5
Main Line Mass
R
P
7
Main Line Flowrate
R
P
9
Process Line Mass
R
P
11
Process Line Flowrate
R
P
13
Process Mass Ratio
R
P
15
Process Flowrate Ratio
R
P
17
Process Control Output
R
P
19
Process Flowrate Deviation
R
P
21
Preset Quantity
R
P
23
R
P
25
R
P
27
R
P
29
R
P
31
Year
Current Date/Time or
Logged Date/Time Stamp
(see register 38 Log Number).
Only current Date/Time can be edited
R/W
I
32
Month
R/W
I
33
Date
R/W
I
34
Hour
R/W
I
35
Minute
R/W
I
36
Second
R
I
37
Log Type
00 - hourly or log records
01 - daily
02 - weekly
03 - monthly
04 - yearly
05 - last edit of calibration
06 -current totals are non-accumulated values,
register 38 is ignored.
R/W
I
38
Log Number
If set to 0, current variables and Date/Time are
retrieved
R/W
I
39
Clear Data
01 - clear logs
02 - clear accumulated totals
03 - clear non-accumulated totals
Note:
the Clear Data command is not executed
(and register 39 is set to non-zero value) when
the controller has a transaction in progress.
R/W
I
40
Reserved