Installation & Maintenance
20
515 BR03 - 15 August 2021
Battery Replacement
INSTRUMENT MUST BE POWERED DOWN
1. Remove the top row of terminal blocks (19-36) and any communication
cables if applicable.
2. Remove the two pan head machine screws on each side of the rear panel.
3. Withdraw the option card from the instrument.
4. Firmly press the battery holder tab to release the coin cell battery.
5. Identify the part number of the coin cell which is clearly marked on one
side of the cell.
6. Firmly press the replacement coin cell into the holder
7. Reverse steps 1-3 for re assembly.
Battery Type
Only the battery type and manufacturer stated below
should be used.
1. Instruments manufactured with issue 7 option card – CR2450N,
Manufacturer RENATA ONLY*
2. Conformal coated “C” version - Type BR2032, manufacturer Panasonic
ONLY.
3. Non Conformally coated versions :- BR2032, CR2032, Sony or
Panasonic ONLY.
*Issue 7 option card can be identified with 6 way (31-36) ORANGE relay
connector.