![Contrec 515 BR03 Operation Manual Download Page 52](http://html1.mh-extra.com/html/contrec/515-br03/515-br03_operation-manual_2658193052.webp)
Instrument Calibration
42
515 BR03 - 15 August 2021
Changing the Instrument Settings
In Calibration Set mode, the display flashes the item that can be changed.
For option settings, the display flashes the complete option. For a numeric
parameter, the display flashes one digit at a time, you can change the value
of the flashing digit as required, then move the flashing cursor to change
another digit.
Note:
When you change the setting of a parameter, the instrument records
the result as soon as you move to another parameter, or exit from the
Calibration Set mode.
Changing Option Settings
When you display an option that can be changed, the entire option flashes
on the display, such as the choices of ODD, EVEN or NONE for the
communications parity bit checking. Press
or
to change the
option. You can “scroll” through the options in either direction to make a
selection as shown below.
Changing Numeric Settings
The display flashes the digit that can be changed.
Press
to select the digit that you wish to change.
Press
or
to increase or decrease the value of the selected digit.
Changing the Decimal Point
To change the position of the decimal point, press
to move the flashing
selection until the decimal point flashes. Press
or
to move the
decimal point to the right or left as required.
Units of Measurement
The calibration of some parameters is based on the units that are defined for
the relevant variables. These units of measurement can been viewed in the
UNITS menu in the Instrument Settings section below.
ODD
EVEN
NONE
ODD
NONE
EVEN
6
789.123
Summary of Contents for 515 BR03
Page 8: ...viii 515 BR03 15 August 2021...
Page 10: ...x 515 BR03 15 August 2021...
Page 16: ...Introduction 6 515 BR03 15 August 2021...
Page 78: ...Instrument Calibration 68 515 BR03 15 August 2021...
Page 100: ...90 515 BR03 15 August 2021...
Page 106: ...96 515 BR03 15 August 2021...