Operating Instructions
LASSTEC
Weighing System 0521
BAL0521-0001-EN
www.conductix.com
translated document
page 59 of 123
5.3.8.2.5
Temperature Calibration
The temperature calibration enables the temperature measurement for the Twistlocks. Each Twistlock has to be calibrated individually.
In order to carry out the temperature calibration, the current temperature of the Twistlocks must be known. An infrared thermometer is
suitable for this purpose. No test weight is needed to perform this calibration
1.
After selecting
Temperature calibration
an instruction window
is
displayed. Read the instructions carefully and confirm by clicking
OK.
2.
You are than asked to select the Twistlock you want to calibrate. Select it and confirm by clicking
OK.
ADVICE!
If you are not sure about the number of your selected Twistlock take a look at the
Lasstec monitor. The Twistlock number is displayed in the main screen, next to the
Twistlock weight.