Table 2 

MegaPulse 968-B Metallic A Front Panel Features

 

ITEM

 

NAME

 

FUNCTION

 

VOLTAGE

 Adjust Knob 

Adjust the digital voltage set point in the tester. 
Press the voltage knob to display the voltage set point. This setting will blink for a few seconds on the front meter. 
Turn Clockwise to increase the Voltage Setting Point before pressing 

CHARGE

 button. 

POLARITY

 switch 

Toggles the output pulse polarity from Normal to Reverse, Normal for positive and Reverse for Negative, The pulse will appear on the 
Output jack relative to the return jack 
The polarity switch only operates when the 

CHARGE 

indicator is lit and the voltage on the display meter is less than 180V.   The polarity 

is normal when the 

NOR

 indicator is lit and reverse when the 

REV

 indicator is lid.   

NOR REV 

indicator 

Indicates the state of the Output Polarity switch.  

NOR

 indicates Normal (Positive) position.  

REV

 indicates Reverse (Negative) position. 

CHARGE

 switch 

Starts the charge process of the tester capacitor. 
The 

CHARGE

 indicator will turn off after the 

CHARGE

 switch is pressed, and the 

TRIGGER

 indicator will turn on. 

The charge process will stop after 2 minutes if the 

TRIGGER

 button is not pressed. 

CHARGE

 indicator 

This Yellow indicator is lit to show that pressing the 

CHARGE

 switch is the next logical step in a test sequence. 

CHARGE

 indicator is lit when the tester is turn ON an after pressing 

TRIGGER

 button. 

CHARGE

 indicator will go out after pressing 

CHARGE

 button. 

CHARGE

 and 

TRIGGER

 Indicators will be blinking if the Interlock Switch is open. (Only testers with Interlock Switch Option) 

TRIGGER

 switch 

Triggers the output impulse waveform. 
The impulse waveform will appear across the output leads. 

TRIGGER

 indicator 

This Red indicator is lit to show that the tester can be triggered. 

TRIGGER

 indicator is lit for 2 minutes after the 

CHARGE

 button is pressed. 

TRIGGER

 indicator will go out after pressing 

TRIGGER

 button. 

 

VOLTAGE

 meter 

Displays the output voltage set point. 
The voltage reading will increase from zero to the voltage set point when the CHARGE button is pressed.  
The meter will start to flash at 

900V

 to indicate that voltage is in the maximum limits. 

If unit includes PC Interface and the Keyboard is locked, the display will show OFF when a button is pressed. 

OUTPUT

 jack 

The impulse waveform appears on the 

OUTPUT

 jack, referenced to the 

RETURN 

jack. When the 

POLARITY

 switch is in the Normal 

position (

NOR

 indicator is lit) the output will be a positive pulse.  When the 

POLARITY

 switch is in the Reverse position (

REV

 indicator 

is lit) the output will be a negative pulse. 

10 

RETURN

 jack 

This is the return for the impulse waveform.  This jack is referenced to the chassis of the MegaPulse and is referenced to earth ground as 
long as the MegaPulse is properly grounded.  Even though this jack is referenced to ground, it should be treated as hazardous whenever 
the MegaPulse is turned ON. 

Summary of Contents for MegaPulse 968-B Metallic A

Page 1: ...Instruction Manual...

Page 2: ...i...

Page 3: ...gn and ergonomic front panel and represents the latest in high voltage impulse testing To fully appreciate all the features of your new instrument we suggest that you take a few moments to review this...

Page 4: ...iii...

Page 5: ...ng the Instrument 1 AC Line Voltage Requirements 2 Fuse Replacement 2 Section 3 3 Specifications and Controls 3 Front and Rear Panel Features 4 Section 4 7 Operating Instructions 7 External Interlock...

Page 6: ...v...

Page 7: ...nclude a warning against any metal jewelry Operators should not allow others in the testing area especially when tests are being conducted Organization is to be stressed The operator should keep the a...

Page 8: ...r one touch operation with the right hand If possible it should be set up to the left and in front of the equipment under test The equipment under test should be connected to the MegaPulse tester and...

Page 9: ...pulse relative to the RETURN jack When the polarity switch is in the Reverse position REV indicator is lit the high voltage will appear on both OUTPUTS as a negative pulse relative to the RETURN jack...

Page 10: ...748 6224 Product Package for 968 B Metallic A Tester Description Part Number 968 B Metallic A MegaPulse Tester User Manual High Voltage Test Lead Red 00 HVL7R Qty 1 High Voltage Test Lead Black 00 HV...

Page 11: ...l test house are regarded as meeting this requirement Fuse Replacement There is a user replaceable fuse F1 located on the rear panel of the instrument It is located behind a door in the Power Inlet Po...

Page 12: ...tage Pulse Front Time 6 s to 10 s 1 67 times between 30 and 90 per TIA 968 B Fig 2 Voltage Pulse Decay Time 560 s to 860 s per TIA 968 B Fig 2 Current Peak 100A minimum Current Pulse Front Time 5 s to...

Page 13: ...amiliar with the use of its controls indicators and connectors The front panel features of the MegaPulse are shown in Figure 2 and described in Table 2 The rear panel features of the MegaPulse are sho...

Page 14: ...er is turn ON an after pressing TRIGGER button CHARGE indicator will go out after pressing CHARGE button CHARGE and TRIGGER Indicators will be blinking if the Interlock Switch is open Only testers wit...

Page 15: ...se replacement warning Rating of power supply Specifies replacement fuse and required supply voltage 3 RS 232 Interface Optional Allow the communication between the tester and computer interface a RS...

Page 16: ...wn on the computer condition that will also disable polarity changes charge and trigger conditions When the Interlock is closed it enables all normal operations of the MegaPulse features Front Keyboar...

Page 17: ...Tester is capable to maintain a specific voltage charge selected by the Voltage Set Point CAUTION High voltage generated by the MegaPulse tester is exposed during this test A risk of shock exists Exer...

Page 18: ...the jacks on the front panel 6 Connect the ends of the test leads to the measuring instrument See Error Reference source not found for an example 7 Set the next capturing setting on the digital oscill...

Page 19: ...neglect Malfunctions which occur within the limits of the warranty will be corrected at no charge Mail the instrument post paid to the manufacturer Dated proof of purchase is required for all in warr...

Page 20: ...Pulse but it will not be the same amount it will depend on the waveform calibration and tolerances therefore it is important to measure the peak amplitude of the output waveform and adjust the output...

Page 21: ...m Compliance West USA is available Phone 800 748 6224 Hours 8 30 AM 4 30 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D S...

Reviews: