Turn the rear-panel Power Switch ON. 

4.

 

Press  the  red 

TRIGGER

  button  to  discharge  the  main  capacitor  into  the  internal  resistor  of  the 

MegaPulse unit. 

5.

 

Turn the rear-panel Power Switch OFF. 

 

Voltage Set Point Adjustment 

 

NOTE 

If the front display shows the word “Off” when trying the adjustment, it means the front Voltage 

knob has been disabled by the Testminder software. See section “Front Keyboard and Voltage Knob 

Enable” 

 

To adjust the Voltage Set Point: 
 

1.

 

Press the VOLTAGE Adjust knob one time to enable the adjustment mode, digits for thousands and 
hundreds will start blinking for a few seconds.  

2.

 

While the digits are still blinking, turn the VOLTAGE Adjust knob, clockwise or counterclockwise, 
to change the value of the blinking digits. 

3.

 

While digits are still blinking, press the VOLTAGE Adjust knob again and the tens and unit digits 
will start blinking for a few seconds.  

4.

 

While the digits are blinking, turn the VOLTAGE Adjust knob, clockwise or counterclockwise, to 
change the value of the blinking digits. 

5.

 

After a few seconds without any change on the VOLTAGE Adjust knob, the desired output voltage 
will be set.  

 

Charge and Trigger a Pulse 

 
The following procedure  will explain how to generate a high  voltage pulse  with the MegaPulse  unit. The 
MegaPulse 968-B Metallic A Tester is capable to maintain a specific voltage charge selected by the Voltage 
Set Point. 

CAUTION

 

High voltage generated by the MegaPulse tester is exposed during this test. A risk of shock exists.  
Exercise care when using the MegaPulse unit. 

 

1.

 

Confirm the Voltage Set Point by pressing the VOLTAGE  Adjust Knob, the  current  setting  will 
blink for a few seconds on the front meter. 

2.

 

If a different Voltage Set Point is needed it, refer to section “Voltage Set Point Adjustment.”  

3.

 

Push the yellow CHARGE button to start charging the internal high voltage capacitor and wait until 
the front meter reaches value set on Steps 1 or 2. Verify that the red TRIGGER indicator is now lit.  

4.

 

Once the desired  voltage is reached, press the red 

TRIGGER

 button to deliver the  high voltage 

pulse (be sure not to touch the output and return leads when pressing the trigger switch). 

5.

 

Turn the rear-panel Power Switch OFF. 

 

Pulse Verification Procedure 

 

The following procedure will verify that the high voltage pulse is properly generated by the MegaPulse tester. 
We recommend that this procedure be conducted periodically to ensure proper operation of the tester. 
The following items are needed to conduct this procedure:  
 

 

High voltage oscilloscope probe (1000:1) 

 

Digital Oscilloscope 

 

Always ensure that the measuring instruments are rated for the voltage and frequency response involved.  A 
digital oscilloscope or probe with a low frequency response will result in erroneous readings that could be 
misread.  

Summary of Contents for MegaPulse 968-B Metallic A

Page 1: ...Instruction Manual...

Page 2: ...i...

Page 3: ...gn and ergonomic front panel and represents the latest in high voltage impulse testing To fully appreciate all the features of your new instrument we suggest that you take a few moments to review this...

Page 4: ...iii...

Page 5: ...ng the Instrument 1 AC Line Voltage Requirements 2 Fuse Replacement 2 Section 3 3 Specifications and Controls 3 Front and Rear Panel Features 4 Section 4 7 Operating Instructions 7 External Interlock...

Page 6: ...v...

Page 7: ...nclude a warning against any metal jewelry Operators should not allow others in the testing area especially when tests are being conducted Organization is to be stressed The operator should keep the a...

Page 8: ...r one touch operation with the right hand If possible it should be set up to the left and in front of the equipment under test The equipment under test should be connected to the MegaPulse tester and...

Page 9: ...pulse relative to the RETURN jack When the polarity switch is in the Reverse position REV indicator is lit the high voltage will appear on both OUTPUTS as a negative pulse relative to the RETURN jack...

Page 10: ...748 6224 Product Package for 968 B Metallic A Tester Description Part Number 968 B Metallic A MegaPulse Tester User Manual High Voltage Test Lead Red 00 HVL7R Qty 1 High Voltage Test Lead Black 00 HV...

Page 11: ...l test house are regarded as meeting this requirement Fuse Replacement There is a user replaceable fuse F1 located on the rear panel of the instrument It is located behind a door in the Power Inlet Po...

Page 12: ...tage Pulse Front Time 6 s to 10 s 1 67 times between 30 and 90 per TIA 968 B Fig 2 Voltage Pulse Decay Time 560 s to 860 s per TIA 968 B Fig 2 Current Peak 100A minimum Current Pulse Front Time 5 s to...

Page 13: ...amiliar with the use of its controls indicators and connectors The front panel features of the MegaPulse are shown in Figure 2 and described in Table 2 The rear panel features of the MegaPulse are sho...

Page 14: ...er is turn ON an after pressing TRIGGER button CHARGE indicator will go out after pressing CHARGE button CHARGE and TRIGGER Indicators will be blinking if the Interlock Switch is open Only testers wit...

Page 15: ...se replacement warning Rating of power supply Specifies replacement fuse and required supply voltage 3 RS 232 Interface Optional Allow the communication between the tester and computer interface a RS...

Page 16: ...wn on the computer condition that will also disable polarity changes charge and trigger conditions When the Interlock is closed it enables all normal operations of the MegaPulse features Front Keyboar...

Page 17: ...Tester is capable to maintain a specific voltage charge selected by the Voltage Set Point CAUTION High voltage generated by the MegaPulse tester is exposed during this test A risk of shock exists Exer...

Page 18: ...the jacks on the front panel 6 Connect the ends of the test leads to the measuring instrument See Error Reference source not found for an example 7 Set the next capturing setting on the digital oscill...

Page 19: ...neglect Malfunctions which occur within the limits of the warranty will be corrected at no charge Mail the instrument post paid to the manufacturer Dated proof of purchase is required for all in warr...

Page 20: ...Pulse but it will not be the same amount it will depend on the waveform calibration and tolerances therefore it is important to measure the peak amplitude of the output waveform and adjust the output...

Page 21: ...m Compliance West USA is available Phone 800 748 6224 Hours 8 30 AM 4 30 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D S...

Reviews: