background image

 

 

Section 1 

An Introduction to Impulse Testing with the MegaPulse series tester 

 

The impulse test is designed to simulate impulse surges which occur in everyday life due to nearby lightning 
strikes,  switching  transients,  and  other  high-frequency  faults  on  the  power  distribution  network.    Impulse 
testing is the fundamental method for empirical verification of the adequacy of insulation. Other methods of 
ensuring adequate insulation (AC or DC Dielectric Withstand testing, measurement of over-surface creepage, 
through-air clearance, or distance-through-insulation) are all extrapolated from the results of impulse testing. 
The impulse test is performed to ensure that the insulation in question will be able to function properly when 
subjected to similar impulse surges in the field.  
 

Safety Precautions 

 
The impulse withstand test can generate voltages in excess of 800V peak at potentially lethal current levels.  
Currents of as little as 5 mA at 120 volts can cause death; the MegaPulse can deliver currents of more than 
100 Amps peak for a very short time duration. The potential for serious injury or death exists and personnel 
should be aware when they conduct this test. 
 

Test Personnel 

 
Personnel  require  special  training  to  conduct  the  impulse  test.  They  should  understand  electrical 
fundamentals clearly and be aware that high voltage is adept and creative at completing a path to ground.  
Instructions should include a warning against any metal jewelry. Operators should not allow others in the 
testing area, especially when tests are being conducted.  Organization is to be stressed. The operator should 
keep the area free of unused leads and equipment. 
 

Testing Area 

 
The area used for conducting the impulse test should be as remote as possible from normal production line 
activities. Only personnel conducting the test should be allowed in the area, and it should be taped or roped 
off to preclude casual entry by other employees. In addition, the area should be marked "WARNING - HIGH 
VOLTAGE TESTING" or the equivalent to warn others of the nature of the testing taking place. 
 
The bench being used should be non-conductive, and any exposed metal parts should be tied together and 
grounded. If a conductive surface must be used, it should be grounded. Because of sparking during an impulse 
test failure, it is not safe to conduct impulse testing in combustible atmospheres. 
 
It is imperative that a good ground be provided to the MegaPulse tester. Before connecting the equipment, 
ensure that the building wiring provides a low-resistance ground. If the MegaPulse tester is used on a high-
resistance grounding circuit, dangerous high voltages may be present to the operator. In addition, the power 
to  the  Testing  Area  should  be  provided  with  an  easily  reached  shutoff  switch  which  can  be  actuated  by 
personnel outside the Area if needed. 
 

CAUTION:

 Plug on power cord is disconnect device. The power to the Testing Area should be provided 

with an 

easily reached shutoff switch

 which can be actuated by personnel outside the area if needed. 

 

 

Summary of Contents for MegaPulse 968-B Metallic A

Page 1: ...Instruction Manual...

Page 2: ...i...

Page 3: ...gn and ergonomic front panel and represents the latest in high voltage impulse testing To fully appreciate all the features of your new instrument we suggest that you take a few moments to review this...

Page 4: ...iii...

Page 5: ...ng the Instrument 1 AC Line Voltage Requirements 2 Fuse Replacement 2 Section 3 3 Specifications and Controls 3 Front and Rear Panel Features 4 Section 4 7 Operating Instructions 7 External Interlock...

Page 6: ...v...

Page 7: ...nclude a warning against any metal jewelry Operators should not allow others in the testing area especially when tests are being conducted Organization is to be stressed The operator should keep the a...

Page 8: ...r one touch operation with the right hand If possible it should be set up to the left and in front of the equipment under test The equipment under test should be connected to the MegaPulse tester and...

Page 9: ...pulse relative to the RETURN jack When the polarity switch is in the Reverse position REV indicator is lit the high voltage will appear on both OUTPUTS as a negative pulse relative to the RETURN jack...

Page 10: ...748 6224 Product Package for 968 B Metallic A Tester Description Part Number 968 B Metallic A MegaPulse Tester User Manual High Voltage Test Lead Red 00 HVL7R Qty 1 High Voltage Test Lead Black 00 HV...

Page 11: ...l test house are regarded as meeting this requirement Fuse Replacement There is a user replaceable fuse F1 located on the rear panel of the instrument It is located behind a door in the Power Inlet Po...

Page 12: ...tage Pulse Front Time 6 s to 10 s 1 67 times between 30 and 90 per TIA 968 B Fig 2 Voltage Pulse Decay Time 560 s to 860 s per TIA 968 B Fig 2 Current Peak 100A minimum Current Pulse Front Time 5 s to...

Page 13: ...amiliar with the use of its controls indicators and connectors The front panel features of the MegaPulse are shown in Figure 2 and described in Table 2 The rear panel features of the MegaPulse are sho...

Page 14: ...er is turn ON an after pressing TRIGGER button CHARGE indicator will go out after pressing CHARGE button CHARGE and TRIGGER Indicators will be blinking if the Interlock Switch is open Only testers wit...

Page 15: ...se replacement warning Rating of power supply Specifies replacement fuse and required supply voltage 3 RS 232 Interface Optional Allow the communication between the tester and computer interface a RS...

Page 16: ...wn on the computer condition that will also disable polarity changes charge and trigger conditions When the Interlock is closed it enables all normal operations of the MegaPulse features Front Keyboar...

Page 17: ...Tester is capable to maintain a specific voltage charge selected by the Voltage Set Point CAUTION High voltage generated by the MegaPulse tester is exposed during this test A risk of shock exists Exer...

Page 18: ...the jacks on the front panel 6 Connect the ends of the test leads to the measuring instrument See Error Reference source not found for an example 7 Set the next capturing setting on the digital oscill...

Page 19: ...neglect Malfunctions which occur within the limits of the warranty will be corrected at no charge Mail the instrument post paid to the manufacturer Dated proof of purchase is required for all in warr...

Page 20: ...Pulse but it will not be the same amount it will depend on the waveform calibration and tolerances therefore it is important to measure the peak amplitude of the output waveform and adjust the output...

Page 21: ...m Compliance West USA is available Phone 800 748 6224 Hours 8 30 AM 4 30 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D S...

Reviews: