background image

 

 

11 

 

1.

 

Connect the Tester to a proper source of supply using the included 18 AWG power supply cord.   

2.

 

Plug the Output and Return test leads in to the jacks on the front panel. 

3.

 

Connect the ends of the test leads to an oscilloscope with a high-voltage probe. Note that the 

RETURN 

lead is referenced to the chassis of the MegaPulse tester.  

4.

 

Turn the Tester on. Toggle the 

POLARITY

 switch if necessary so that the 

NOR

 indicator is lit. 

5.

 

Note  that  the  Voltage  meter  may  indicate  that  some  residual  voltage  is  present  on  the  main  storage 
capacitor,  even  when  the  tester  is  first  turned  ON.  This  is  due  to  inherent  charging  of  the  internal 
capacitors. Pressing the 

TRIGGER

 switch will discharge the capacitors (be sure not to touch the outputs 

and return leads when pressing the trigger switch). 

6.

 

Adjust the 

VOLTAGE

 using the voltage knob so that the front panel LED display is reading a voltage 

that is suitable for the measuring instrument that is being used - preferably close to the maximum output 
voltage of the MegaPulse tester.  

7.

 

Press the 

CHARGE

 button and, when the desired voltage is reached, push the red 

TRIGGER

 button, 

and view the resulting impulse waveform on the measuring instrument. 

8.

 

Evaluate the voltage waveforms (peak, rise and duration) that you captured, they should be within the 
tolerances  mentioned  in 

Table  1

.  If  the  measured  waveforms  are  out  of  tolerances,  then  additional 

evaluation and/or repair of the MegaPulse tester is needed.  

9.

 

Note that the peak amplitude of the measured output waveform is proportional to the voltage that is read 
of the front panel of the MegaPulse, but it will not be the same amount, it will depend on the waveform 
calibration  and  tolerances,  therefore,  it  is  important  to  measure  the  peak  amplitude  of  the  output 
waveform and adjust the output of the MegaPulse accordingly. 

10.

 

Repeat steps 5 through 9, except this time toggle the 

POLARITY

 switch so that the 

REV

 indicator is 

lit. Note that the impulse  waveform will now be a negative pulse, so it will probably be necessary to 
adjust the measuring instrument to get a proper reading.   

11.

 

Connect the Output and the Return test leads across an appropriate current shunt, so that the Output lead 
is connected to one end of the current shunt, and the Return lead is connected to the other end of the 
current shunt. 

12.

 

Connect  the  measuring  instrument  across  the  current  shunt,  keeping  in  mind  that  the  Return  lead  is 
referenced to ground. 

13.

 

Toggle the 

POLARITY

 switch if necessary so that the 

NOR

 indicator is lit. 

14.

 

Press the 

CHARGE

 button and, when the desired voltage is reached, push the red 

TRIGGER

 button, 

and view the resulting impulse waveform on the measuring instrument. 

15.

 

Evaluate the current waveforms (peak, rise and duration) that you  captured, they should be within the 
tolerances  mentioned  in 

Table  1

.  If  the  measured  waveforms  are  out  of  tolerances,  then  additional 

evaluation and/or repair of the MegaPulse tester is needed. 

16.

 

Repeat steps 14 through 17, except this time toggle the 

POLARITY

 switch so that the 

REV

 indicator 

is lit. Note that the impulse waveform will now be a negative pulse, so it will probably be necessary to 
adjust the measuring instrument to get a proper reading. 

17.

 

Turn the rear-panel power switch OFF. 

 
If the waveforms are out of tolerances, service is required. Remove the Tester from service and contact the 
manufacturer for servicing information.

Summary of Contents for MegaPulse 968-B Metallic A

Page 1: ...Instruction Manual...

Page 2: ...i...

Page 3: ...gn and ergonomic front panel and represents the latest in high voltage impulse testing To fully appreciate all the features of your new instrument we suggest that you take a few moments to review this...

Page 4: ...iii...

Page 5: ...ng the Instrument 1 AC Line Voltage Requirements 2 Fuse Replacement 2 Section 3 3 Specifications and Controls 3 Front and Rear Panel Features 4 Section 4 7 Operating Instructions 7 External Interlock...

Page 6: ...v...

Page 7: ...nclude a warning against any metal jewelry Operators should not allow others in the testing area especially when tests are being conducted Organization is to be stressed The operator should keep the a...

Page 8: ...r one touch operation with the right hand If possible it should be set up to the left and in front of the equipment under test The equipment under test should be connected to the MegaPulse tester and...

Page 9: ...pulse relative to the RETURN jack When the polarity switch is in the Reverse position REV indicator is lit the high voltage will appear on both OUTPUTS as a negative pulse relative to the RETURN jack...

Page 10: ...748 6224 Product Package for 968 B Metallic A Tester Description Part Number 968 B Metallic A MegaPulse Tester User Manual High Voltage Test Lead Red 00 HVL7R Qty 1 High Voltage Test Lead Black 00 HV...

Page 11: ...l test house are regarded as meeting this requirement Fuse Replacement There is a user replaceable fuse F1 located on the rear panel of the instrument It is located behind a door in the Power Inlet Po...

Page 12: ...tage Pulse Front Time 6 s to 10 s 1 67 times between 30 and 90 per TIA 968 B Fig 2 Voltage Pulse Decay Time 560 s to 860 s per TIA 968 B Fig 2 Current Peak 100A minimum Current Pulse Front Time 5 s to...

Page 13: ...amiliar with the use of its controls indicators and connectors The front panel features of the MegaPulse are shown in Figure 2 and described in Table 2 The rear panel features of the MegaPulse are sho...

Page 14: ...er is turn ON an after pressing TRIGGER button CHARGE indicator will go out after pressing CHARGE button CHARGE and TRIGGER Indicators will be blinking if the Interlock Switch is open Only testers wit...

Page 15: ...se replacement warning Rating of power supply Specifies replacement fuse and required supply voltage 3 RS 232 Interface Optional Allow the communication between the tester and computer interface a RS...

Page 16: ...wn on the computer condition that will also disable polarity changes charge and trigger conditions When the Interlock is closed it enables all normal operations of the MegaPulse features Front Keyboar...

Page 17: ...Tester is capable to maintain a specific voltage charge selected by the Voltage Set Point CAUTION High voltage generated by the MegaPulse tester is exposed during this test A risk of shock exists Exer...

Page 18: ...the jacks on the front panel 6 Connect the ends of the test leads to the measuring instrument See Error Reference source not found for an example 7 Set the next capturing setting on the digital oscill...

Page 19: ...neglect Malfunctions which occur within the limits of the warranty will be corrected at no charge Mail the instrument post paid to the manufacturer Dated proof of purchase is required for all in warr...

Page 20: ...Pulse but it will not be the same amount it will depend on the waveform calibration and tolerances therefore it is important to measure the peak amplitude of the output waveform and adjust the output...

Page 21: ...m Compliance West USA is available Phone 800 748 6224 Hours 8 30 AM 4 30 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D S...

Reviews: