Section 1 

 

Safety Precautions 

 
 

The  dielectric  withstand  test  generates  voltages  of  up  to  3000  volts  ac  at  potentially 

lethal  current  levels.  Currents  of  as  little  as  5  mA  at  120  volts  can  cause  death,  and  the  HT-
3000P-1A-AC is capable of generating 1 A at up to 3000 volts. The HT-3000P-1A-AC has been 
designed  to  minimize  exposure  to  high  voltages.  However,  the  potential  for  serious  injury  or 
death exists and personnel should be aware when they conduct this test. 

 

Test Personnel 

 
 

Personnel  require  special  training  to  conduct  the  dielectric  withstand  test.  They  should 

understand electrical fundamentals clearly, and be aware that high voltage is adept and creative 
at completing a path to ground. Instructions should include a warning against any metal jewelry.  
Operators should not allow others in the testing area, especially when tests are being conducted.  
Organization  is  to  be  stressed.  The  operator  should  keep  the  area  free  of  unused  leads  and 
equipment. 

 

Testing Area 

 

 

The area used for conducting the dielectric withstand test should be as remote as possible 

from  normal  production  line  activities.  Only  personnel  actually  conducting  the  test  should  be 
allowed  in  the  area,  and  it  should  be  taped  or  roped  off  to  preclude  casual  entry  by  other 
employees. In addition, the area should be marked "WARNING - HIGH VOLTAGE TESTING" 
or the equivalent to warn others of the nature of the testing taking place.   
 

The bench being used should be non-conductive, and any exposed metal parts should be 

tied together and grounded.  If a conductive surface must be used, it should be grounded. 
 

Because  of  sparking  during  a  dielectric  test  failure,  it  is  not  safe  to  conduct  dielectric 

withstand tests in combustible atmospheres. 
 

It  is  imperative  that  a  good  ground  be  provided  to  the  HT-3000P-1A-AC.  Before 

connecting  the  HT-3000P-1A-AC,  ensure  that  a  low-resistance  ground  is  provided  by  the 
building  wiring.  If  the  HT-3000P-1A-AC  is  used  on  a  high-resistance  grounding  circuit, 
dangerous  high  voltages  may  be  present  to  the  operator.  In  addition,  the  power  to  the  Testing 
Area  should  be  provided  with  an  easily  reached  shutoff  switch  which  can  be  actuated  by 
personnel outside the Area if needed. 

Summary of Contents for HT-3000P-1A-AC

Page 1: ...HT 3000P 1A AC Dielectric Withstand Tester 0 3000 Volts AC Output Instruction Manual...

Page 2: ......

Page 3: ......

Page 4: ...micro controlled circuit design and ergonomic front panel and represents the latest in high voltage production line testing To fully appreciate all the features of your new meter we suggest that you t...

Page 5: ...ii...

Page 6: ...al Checkout Procedure 10 Performance Test 10 Setting up the HT 3000P 1A AC 10 Factory Settings 11 Display of Leakage Limit and Duration settings 11 Adjustment of the Leakage Current Level 11 Adjustmen...

Page 7: ...1...

Page 8: ...for conducting the dielectric withstand test should be as remote as possible from normal production line activities Only personnel actually conducting the test should be allowed in the area and it sho...

Page 9: ...er s return lead is connected to ground potential and when properly connected to the equipment being tested it will guard against the operator contacting high voltage Always make sure the return lead...

Page 10: ...being tested are still failing the primary circuit capacitance of the equipment may be too high to allow the AC dielectric withstand test to be used This is due to the AC voltage charging all capacit...

Page 11: ...sing the RESET Button The test will not begin until the TEST Button is pushed If a failure is encountered the high voltage is immediately shut down a buzzer sounds and any voltage stored in the equipm...

Page 12: ...s Tamperproof operation available ENVIRONMENTAL Operating Temperature 15 40 C Relative Humidity Range 0 90 non condensing GENERAL Input power requirements 240V 60 Hz 15A single phase Weight 180 lbs ap...

Page 13: ...e during shipment We recommend that the instrument be surrounded by at least three inches of shock absorbing material on all sides of the container AC Line Voltage Requirements AC line voltage require...

Page 14: ...8 Figure 2 Controls Indicators Connectors Model HT 3000P 1A AC Front Panel...

Page 15: ...visual indication to the operator of the actual output voltage 9 Hipot Pass LED At the end of the test if no Excess leakage was detected the green hipot pass led will light and the test will terminate...

Page 16: ...y verify the correct leakage current setting 7 Connect the RETURN lead to the HIGH VOLTAGE TEST lead make a short circuit Keep clear of the wire while the test is in progress 8 Push the red RESET butt...

Page 17: ...to 1000 mA To set the leakage limit follow the steps below 1 Connect the HT 3000P 1A AC to a correctly rated source of supply and turn the power switch to the I or ON position Push the RESET button T...

Page 18: ...ng with testing the HT 3000P 1A AC tester should be set up for production line testing See Section above before continuing CAUTION High voltage is generated by the HT 3000P 1A AC Although the chassis...

Page 19: ...ce West USA is available Phone 800 748 6224 Hours 8 00 AM 4 00 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D San Diego C...

Page 20: ...the calibration procedure is a DMM able to read true rms 0 3000 Vac 1 and a 3kohm 3600 w 1 resistor Service Information The HT 3000P 1A AC is warranted to the original purchaser for a period of 1 year...

Page 21: ...Calibration adjustment calibrates the voltage output to agree with the meter reading 2 Leakage Current calibration adjustment calibrates the internal current meter to obtain the real current flowing b...

Page 22: ...you can A Press Reset to exit out of the Calibration Mode and keep all of the currently programmed calibration settings or B Press Test to enter the Calibration Mode and create new calibration settin...

Page 23: ...Return jack on the front panel of the HT 3000P 1A AC 4 In the calibration mode hit the Reset button until L1 is displayed on the front panel 5 Once L1 is displayed press the Test button once and 80 w...

Reviews: