Adjustment of the High Voltage Test Time 

 
This procedure sets the length of time the HT-3000P-1A-AC will conduct the high voltage 

test.  Consult  the  safety  agencies  for  the  test  time  for  the  type  of  equipment  being  tested.  If  a 
different test time is required, use this procedure to set it.   
 
 
1. 

Connect  the  HT-3000P-1A-AC  to  a  correctly  rated  source  of  supply  and  turn  the  power 
switch to the I or ON position. Push the RESET button. The yellow TEST indicator should 
light, indicating that the HT-3000P-1A-AC is ready to test. 

2. 

Turn the Test Time potentiometer.  As soon as the potentiometer starts turning, the meter 
will display “d” and the value can be set in 1 second increments from 1 to 60 seconds.

 

 

Operating Techniques 

 

The  following  paragraphs  describe  how  to  operate  your  HT-3000P-1A-AC  Dielectric 

Withstand Tester. Before proceeding with testing, the HT-3000P-1A-AC tester should be set up 
for production line testing. See Section above before continuing. 
 

CAUTION: 

High  voltage  is  generated  by  the  HT-3000P-1A-AC.  Although  the  chassis  of  the 

equipment under test is grounded by the HT-3000P-1A-AC, a risk of shock exists. Exercise 
care when using the HT-3000P-1A-AC. 

 

Daily Operation Test 

 
The operation of the  HT-3000P-1A-AC should be audited daily by conducting the tests 

described in the Initial Checkout Procedure section of this Manual. 
 

Test results 

 

1.

 

Excess leakage LED/Buzzer 

 

It means the equipment being tested failed. The equipment should be set aside, reworked and 

retested  with acceptable results before it is shipped.  General interpretations of test failures are 
included in Section 1, "Using the HT-3000P-1A-AC Dielectric Withstand Tester". 

 
2.

 

Hipot fail LED/Buzzer 

 

It means the equipment has detected a dangerous arc and stops the test. It is a safety feature, 

it  is  recommended  to  review  the connections and the  device under test before  applying voltage 
again.

Summary of Contents for HT-3000P-1A-AC

Page 1: ...HT 3000P 1A AC Dielectric Withstand Tester 0 3000 Volts AC Output Instruction Manual...

Page 2: ......

Page 3: ......

Page 4: ...micro controlled circuit design and ergonomic front panel and represents the latest in high voltage production line testing To fully appreciate all the features of your new meter we suggest that you t...

Page 5: ...ii...

Page 6: ...al Checkout Procedure 10 Performance Test 10 Setting up the HT 3000P 1A AC 10 Factory Settings 11 Display of Leakage Limit and Duration settings 11 Adjustment of the Leakage Current Level 11 Adjustmen...

Page 7: ...1...

Page 8: ...for conducting the dielectric withstand test should be as remote as possible from normal production line activities Only personnel actually conducting the test should be allowed in the area and it sho...

Page 9: ...er s return lead is connected to ground potential and when properly connected to the equipment being tested it will guard against the operator contacting high voltage Always make sure the return lead...

Page 10: ...being tested are still failing the primary circuit capacitance of the equipment may be too high to allow the AC dielectric withstand test to be used This is due to the AC voltage charging all capacit...

Page 11: ...sing the RESET Button The test will not begin until the TEST Button is pushed If a failure is encountered the high voltage is immediately shut down a buzzer sounds and any voltage stored in the equipm...

Page 12: ...s Tamperproof operation available ENVIRONMENTAL Operating Temperature 15 40 C Relative Humidity Range 0 90 non condensing GENERAL Input power requirements 240V 60 Hz 15A single phase Weight 180 lbs ap...

Page 13: ...e during shipment We recommend that the instrument be surrounded by at least three inches of shock absorbing material on all sides of the container AC Line Voltage Requirements AC line voltage require...

Page 14: ...8 Figure 2 Controls Indicators Connectors Model HT 3000P 1A AC Front Panel...

Page 15: ...visual indication to the operator of the actual output voltage 9 Hipot Pass LED At the end of the test if no Excess leakage was detected the green hipot pass led will light and the test will terminate...

Page 16: ...y verify the correct leakage current setting 7 Connect the RETURN lead to the HIGH VOLTAGE TEST lead make a short circuit Keep clear of the wire while the test is in progress 8 Push the red RESET butt...

Page 17: ...to 1000 mA To set the leakage limit follow the steps below 1 Connect the HT 3000P 1A AC to a correctly rated source of supply and turn the power switch to the I or ON position Push the RESET button T...

Page 18: ...ng with testing the HT 3000P 1A AC tester should be set up for production line testing See Section above before continuing CAUTION High voltage is generated by the HT 3000P 1A AC Although the chassis...

Page 19: ...ce West USA is available Phone 800 748 6224 Hours 8 00 AM 4 00 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D San Diego C...

Page 20: ...the calibration procedure is a DMM able to read true rms 0 3000 Vac 1 and a 3kohm 3600 w 1 resistor Service Information The HT 3000P 1A AC is warranted to the original purchaser for a period of 1 year...

Page 21: ...Calibration adjustment calibrates the voltage output to agree with the meter reading 2 Leakage Current calibration adjustment calibrates the internal current meter to obtain the real current flowing b...

Page 22: ...you can A Press Reset to exit out of the Calibration Mode and keep all of the currently programmed calibration settings or B Press Test to enter the Calibration Mode and create new calibration settin...

Page 23: ...Return jack on the front panel of the HT 3000P 1A AC 4 In the calibration mode hit the Reset button until L1 is displayed on the front panel 5 Once L1 is displayed press the Test button once and 80 w...

Reviews: