12.  Turn the Voltage Adjust on the rear panel clockwise until the output on the DMM equals 

500  volts  AC  and  press  the  Test  button.    The  display  will  now  display  "V2"  again.    The 
voltage is now be calibrated. 

13.  Turn Off the HT-3000P-1A-AC tester. 
14.  The voltage can be verified by turning on the HT-3000P-1A-AC, running a test, and using 

the  1000:1  voltage  probe  to  check  the  meter  at  different  voltages.    If the voltages are not 
/- 2% +/- 1 count at each voltage, the voltage calibration must be performed again 
starting at step 1 above. 

 

Current Calibration Adjustment 

 
1. 

You must be in Calibration Mode as in step 3 under the Voltage Calibration Adjustment 
above. 

2. 

Turn the Voltage Adjust on the rear panel to minimum (counterclockwise).  

3. 

Connect the output of the HT-3000P-1A-AC through a 3K ohm, 3600 watt high voltage 
resistor in series with a calibrated DMM capable of measuring 1A AC and returning to the 
Return jack on the front panel of the HT-3000P-1A-AC. 

4. 

In the calibration mode, hit the Reset button until "L1" is displayed on the front panel. 

5. 

Once "L1" is displayed, press the Test button once and .80 will be displayed. 

6. 

Slowly increase the variac watching the current flowing on the DMM.  Adjust the variac 
until 0.80 A AC is displayed on the DMM and then press the Test button. 

7. 

The display will show "hold". 

8. 

Wait until the display returns to "L1". 

9. 

Press Reset button and meter will display “L2”. Press the Test button once and .10 will be 
displayed. 

10. 

Slowly increase the variac watching the current flowing on the DMM, Adjust the variac 
until 0.10 A AC is displayed on the DMM and then press the Test button. 

11. 

The display will show “hold”. 

12. 

Wait until the display returns to “L2”. At this point the leakage current is calibrated. 

13.  The current can be verified by turning on the HT-3000P-1A-AC, running a test, and using 

DMM in series with the load measuring current at different current levels.  If the current is 
not /-  2%  +/- 1 count, the current calibration must be performed again starting at 
step 1 above. 

 

Summary of Contents for HT-3000P-1A-AC

Page 1: ...HT 3000P 1A AC Dielectric Withstand Tester 0 3000 Volts AC Output Instruction Manual...

Page 2: ......

Page 3: ......

Page 4: ...micro controlled circuit design and ergonomic front panel and represents the latest in high voltage production line testing To fully appreciate all the features of your new meter we suggest that you t...

Page 5: ...ii...

Page 6: ...al Checkout Procedure 10 Performance Test 10 Setting up the HT 3000P 1A AC 10 Factory Settings 11 Display of Leakage Limit and Duration settings 11 Adjustment of the Leakage Current Level 11 Adjustmen...

Page 7: ...1...

Page 8: ...for conducting the dielectric withstand test should be as remote as possible from normal production line activities Only personnel actually conducting the test should be allowed in the area and it sho...

Page 9: ...er s return lead is connected to ground potential and when properly connected to the equipment being tested it will guard against the operator contacting high voltage Always make sure the return lead...

Page 10: ...being tested are still failing the primary circuit capacitance of the equipment may be too high to allow the AC dielectric withstand test to be used This is due to the AC voltage charging all capacit...

Page 11: ...sing the RESET Button The test will not begin until the TEST Button is pushed If a failure is encountered the high voltage is immediately shut down a buzzer sounds and any voltage stored in the equipm...

Page 12: ...s Tamperproof operation available ENVIRONMENTAL Operating Temperature 15 40 C Relative Humidity Range 0 90 non condensing GENERAL Input power requirements 240V 60 Hz 15A single phase Weight 180 lbs ap...

Page 13: ...e during shipment We recommend that the instrument be surrounded by at least three inches of shock absorbing material on all sides of the container AC Line Voltage Requirements AC line voltage require...

Page 14: ...8 Figure 2 Controls Indicators Connectors Model HT 3000P 1A AC Front Panel...

Page 15: ...visual indication to the operator of the actual output voltage 9 Hipot Pass LED At the end of the test if no Excess leakage was detected the green hipot pass led will light and the test will terminate...

Page 16: ...y verify the correct leakage current setting 7 Connect the RETURN lead to the HIGH VOLTAGE TEST lead make a short circuit Keep clear of the wire while the test is in progress 8 Push the red RESET butt...

Page 17: ...to 1000 mA To set the leakage limit follow the steps below 1 Connect the HT 3000P 1A AC to a correctly rated source of supply and turn the power switch to the I or ON position Push the RESET button T...

Page 18: ...ng with testing the HT 3000P 1A AC tester should be set up for production line testing See Section above before continuing CAUTION High voltage is generated by the HT 3000P 1A AC Although the chassis...

Page 19: ...ce West USA is available Phone 800 748 6224 Hours 8 00 AM 4 00 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D San Diego C...

Page 20: ...the calibration procedure is a DMM able to read true rms 0 3000 Vac 1 and a 3kohm 3600 w 1 resistor Service Information The HT 3000P 1A AC is warranted to the original purchaser for a period of 1 year...

Page 21: ...Calibration adjustment calibrates the voltage output to agree with the meter reading 2 Leakage Current calibration adjustment calibrates the internal current meter to obtain the real current flowing b...

Page 22: ...you can A Press Reset to exit out of the Calibration Mode and keep all of the currently programmed calibration settings or B Press Test to enter the Calibration Mode and create new calibration settin...

Page 23: ...Return jack on the front panel of the HT 3000P 1A AC 4 In the calibration mode hit the Reset button until L1 is displayed on the front panel 5 Once L1 is displayed press the Test button once and 80 w...

Reviews: