14 

Section 5 

Maintenance and Calibration 

 
 

WARNING 

THESE  SERVICE  INSTRUCTIONS  ARE  FOR  USE  BY  QUALIFIED  PERSONNEL 
ONLY.  TO  AVOID  ELECTRIC  SHOCK,  DO  NOT  PERFORM  ANY  SERVICING 
OTHER  THAN  THAT  CONTAINED  IN  THE  OPERATING  INSTRUCTIONS  UNLESS 
YOU ARE QUALIFIED TO DO SO. 
 

Introduction 

 

This section of the manual contains maintenance information for the HT-3000P-1A-AC 

Dielectric  Withstand  Tester.  This  maintenance  information  is  divided  into  service  information, 
general  maintenance,  a  performance  test,  and  a  calibration  procedure.  The  performance  test  is 
recommended  as  an  acceptance  test  when  the  instrument  is  first  received,  and  later  as  a 
preventative maintenance tool to verify proper instrument operation. A 1-year calibration cycle is 
recommended to maintain the specifications given in Section 1. The test equipment required for 
the calibration procedure is a DMM able to read true rms 0-3000 Vac 

±

 1% and a 3kohm 3600 w 

1% resistor. 
 

Service Information 

 

The HT-3000P-1A-AC is warranted to the original purchaser for a period of 1 year. This 

warranty does not cover problems due to misuse or neglect.  
 

Malfunctions  which  occur  within  the  limits  of  the  warranty  will  be  corrected  at  no 

charge.  Mail the instrument post paid to the manufacturer. Dated proof of purchase is required 
for all in-warranty repairs.   
 

The  manufacturer  is  also  available  for  calibration  and/or  repair  of  instruments  that  are 

beyond their warranty period. Contact the manufacturer for a cost quotation. Ship the instrument 
and your remittance according to the instructions given by the manufacturer. 
 

General Maintenance 

 

WARNING 

Dangerous  voltages  exist  when  energized.    Exercise  extreme  care  when  working  on  an 
energized circuit.
 
 

Cleaning 

 

CAUTION 

Do  not  use  aromatic  hydrocarbons  or  chlorinated  solvents  for  cleaning.    These  solutions 
will react with the plastic materials used in the instrument. 

Summary of Contents for HT-3000P-1A-AC

Page 1: ...HT 3000P 1A AC Dielectric Withstand Tester 0 3000 Volts AC Output Instruction Manual...

Page 2: ......

Page 3: ......

Page 4: ...micro controlled circuit design and ergonomic front panel and represents the latest in high voltage production line testing To fully appreciate all the features of your new meter we suggest that you t...

Page 5: ...ii...

Page 6: ...al Checkout Procedure 10 Performance Test 10 Setting up the HT 3000P 1A AC 10 Factory Settings 11 Display of Leakage Limit and Duration settings 11 Adjustment of the Leakage Current Level 11 Adjustmen...

Page 7: ...1...

Page 8: ...for conducting the dielectric withstand test should be as remote as possible from normal production line activities Only personnel actually conducting the test should be allowed in the area and it sho...

Page 9: ...er s return lead is connected to ground potential and when properly connected to the equipment being tested it will guard against the operator contacting high voltage Always make sure the return lead...

Page 10: ...being tested are still failing the primary circuit capacitance of the equipment may be too high to allow the AC dielectric withstand test to be used This is due to the AC voltage charging all capacit...

Page 11: ...sing the RESET Button The test will not begin until the TEST Button is pushed If a failure is encountered the high voltage is immediately shut down a buzzer sounds and any voltage stored in the equipm...

Page 12: ...s Tamperproof operation available ENVIRONMENTAL Operating Temperature 15 40 C Relative Humidity Range 0 90 non condensing GENERAL Input power requirements 240V 60 Hz 15A single phase Weight 180 lbs ap...

Page 13: ...e during shipment We recommend that the instrument be surrounded by at least three inches of shock absorbing material on all sides of the container AC Line Voltage Requirements AC line voltage require...

Page 14: ...8 Figure 2 Controls Indicators Connectors Model HT 3000P 1A AC Front Panel...

Page 15: ...visual indication to the operator of the actual output voltage 9 Hipot Pass LED At the end of the test if no Excess leakage was detected the green hipot pass led will light and the test will terminate...

Page 16: ...y verify the correct leakage current setting 7 Connect the RETURN lead to the HIGH VOLTAGE TEST lead make a short circuit Keep clear of the wire while the test is in progress 8 Push the red RESET butt...

Page 17: ...to 1000 mA To set the leakage limit follow the steps below 1 Connect the HT 3000P 1A AC to a correctly rated source of supply and turn the power switch to the I or ON position Push the RESET button T...

Page 18: ...ng with testing the HT 3000P 1A AC tester should be set up for production line testing See Section above before continuing CAUTION High voltage is generated by the HT 3000P 1A AC Although the chassis...

Page 19: ...ce West USA is available Phone 800 748 6224 Hours 8 00 AM 4 00 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D San Diego C...

Page 20: ...the calibration procedure is a DMM able to read true rms 0 3000 Vac 1 and a 3kohm 3600 w 1 resistor Service Information The HT 3000P 1A AC is warranted to the original purchaser for a period of 1 year...

Page 21: ...Calibration adjustment calibrates the voltage output to agree with the meter reading 2 Leakage Current calibration adjustment calibrates the internal current meter to obtain the real current flowing b...

Page 22: ...you can A Press Reset to exit out of the Calibration Mode and keep all of the currently programmed calibration settings or B Press Test to enter the Calibration Mode and create new calibration settin...

Page 23: ...Return jack on the front panel of the HT 3000P 1A AC 4 In the calibration mode hit the Reset button until L1 is displayed on the front panel 5 Once L1 is displayed press the Test button once and 80 w...

Reviews: