6
Specifications
Specifications for the HT-3000P-1A-AC are listed in Table 1.
ELECTRICAL
Output
0-3000 Volts AC
Leakage Current
30-1000 mA
Pass/Fail Criteria:
Leakage Current:
User adjustable 30-1000 mA
Test Time:
User adjustable 1-60 sec.
Pass/Fail Repeatability
±
3%
Meter accuracy
+ 2% from 500-3000 volts AC
Duty cycle
100 %
Discharge Time
Voltage dependent; discharge progress displayed on
front panel meter
Test adjustments
Tamperproof operation available
ENVIRONMENTAL
Operating Temperature
15-40
°
C
Relative Humidity Range
0-90% non-condensing
GENERAL
Input power requirements
240V, 60 Hz, 15A single phase
Weight
180 lbs approximated
SAFETY AGENCY TOPICS
Transformer Output
>500VA
Visual Indication of Voltage Output Provided by
front panel meter, directly connected to high voltage
output
Failure Indication
Audible, provided by internal buzzer
Visual, provided by red LED's on front panel
Test automatically terminated on failure
Leakage Test provided; 50 mA factory set pass/fail
point, user adjustable.
Table 1. HT-3000P-1A-AC Specifications