16 

 

Calibration and Software Version Information 

 

This will allow the user to see the version of the software as well as who performed the last 

calibration. 

 

1. 

Turn off the HT-3000P-1A-AC tester. 

2. 

Hold in the Reset button while turning on the tester. 

3. 

The meter will display 3 items: 

 

A)  The model number of the tester. 

 

B)  The version of the software. 

 

C)  Laboratory number to designate who performed the last calibration: 

 (1= Compliance West USA, 2= another company)

 

 

 

Voltage Calibration Adjustment 

 

Use the following procedure to calibrate the output voltage.  
 
1. 

Ensure that all test leads are removed from the HT-3000P-1A-AC. 

2. 

Turn OFF the HT-3000P-1A-AC tester. 

3. 

To enter the Calibration Mode you MUST follow the following sequence: 

 

-  Push and hold in both the Test and Reset buttons. 

 

-  Turn on the HT-3000P-1A-AC tester. 

 

-  Release the Test button, release the Reset button, press and hold the Reset button, press 
and hold the Test button, release the Test button, and then release the Reset button. 

4. 

If the correct sequence was entered, the display will read "Sure", if not, start again at step 2 
above. 

5. 

While "Sure" is displayed on the screen you can: 
A)    Press  Reset  to  exit  out  of  the  Calibration  Mode  and  keep  all  of  the  currently 
programmed calibration settings or 
B)      Press  Test  to  enter  the  Calibration  Mode  and  create  new  calibration  settings.    (Be 
sure you want to enter the Calibration Mode as this will change the laboratory number so 
it will show the calibration was not performed by Compliance West USA).  

6. 

Once  the  Calibration  Mode  has  been  entered,  the  Reset  button  toggles  between  the  5 
calibration menus: Volt, V1, V2, curr, and bars.  Make sure "V1" is displayed on the meter. 

7. 

Hook up a calibrated 1000:1 voltage probe and a calibrated DMM between output voltage 
leads. 

8. 

Turn the Voltage Adjust on the rear panel to minimum (counterclockwise).  Press the Test 
button and 3000 will be displayed on the front panel meter.  Be careful as the HT-3000P-
1A-AC will be putting out voltage at this point. 

9. 

Turn the Voltage Adjust on the rear panel clockwise until the output on the DMM equals 
3000 volts AC and press the Test button.  The display will now display "V1" again. 

10.  Press the Reset button until "V2" is displayed on the front panel meter. 
11.  Turn the Voltage Adjust on the rear panel to minimum (counterclockwise).  Press the Test 

button and 500 will be displayed on the front panel meter.  Be careful as the HT-3000P-1A-
AC will be putting out voltage at this point. 

Summary of Contents for HT-3000P-1A-AC

Page 1: ...HT 3000P 1A AC Dielectric Withstand Tester 0 3000 Volts AC Output Instruction Manual...

Page 2: ......

Page 3: ......

Page 4: ...micro controlled circuit design and ergonomic front panel and represents the latest in high voltage production line testing To fully appreciate all the features of your new meter we suggest that you t...

Page 5: ...ii...

Page 6: ...al Checkout Procedure 10 Performance Test 10 Setting up the HT 3000P 1A AC 10 Factory Settings 11 Display of Leakage Limit and Duration settings 11 Adjustment of the Leakage Current Level 11 Adjustmen...

Page 7: ...1...

Page 8: ...for conducting the dielectric withstand test should be as remote as possible from normal production line activities Only personnel actually conducting the test should be allowed in the area and it sho...

Page 9: ...er s return lead is connected to ground potential and when properly connected to the equipment being tested it will guard against the operator contacting high voltage Always make sure the return lead...

Page 10: ...being tested are still failing the primary circuit capacitance of the equipment may be too high to allow the AC dielectric withstand test to be used This is due to the AC voltage charging all capacit...

Page 11: ...sing the RESET Button The test will not begin until the TEST Button is pushed If a failure is encountered the high voltage is immediately shut down a buzzer sounds and any voltage stored in the equipm...

Page 12: ...s Tamperproof operation available ENVIRONMENTAL Operating Temperature 15 40 C Relative Humidity Range 0 90 non condensing GENERAL Input power requirements 240V 60 Hz 15A single phase Weight 180 lbs ap...

Page 13: ...e during shipment We recommend that the instrument be surrounded by at least three inches of shock absorbing material on all sides of the container AC Line Voltage Requirements AC line voltage require...

Page 14: ...8 Figure 2 Controls Indicators Connectors Model HT 3000P 1A AC Front Panel...

Page 15: ...visual indication to the operator of the actual output voltage 9 Hipot Pass LED At the end of the test if no Excess leakage was detected the green hipot pass led will light and the test will terminate...

Page 16: ...y verify the correct leakage current setting 7 Connect the RETURN lead to the HIGH VOLTAGE TEST lead make a short circuit Keep clear of the wire while the test is in progress 8 Push the red RESET butt...

Page 17: ...to 1000 mA To set the leakage limit follow the steps below 1 Connect the HT 3000P 1A AC to a correctly rated source of supply and turn the power switch to the I or ON position Push the RESET button T...

Page 18: ...ng with testing the HT 3000P 1A AC tester should be set up for production line testing See Section above before continuing CAUTION High voltage is generated by the HT 3000P 1A AC Although the chassis...

Page 19: ...ce West USA is available Phone 800 748 6224 Hours 8 00 AM 4 00 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D San Diego C...

Page 20: ...the calibration procedure is a DMM able to read true rms 0 3000 Vac 1 and a 3kohm 3600 w 1 resistor Service Information The HT 3000P 1A AC is warranted to the original purchaser for a period of 1 year...

Page 21: ...Calibration adjustment calibrates the voltage output to agree with the meter reading 2 Leakage Current calibration adjustment calibrates the internal current meter to obtain the real current flowing b...

Page 22: ...you can A Press Reset to exit out of the Calibration Mode and keep all of the currently programmed calibration settings or B Press Test to enter the Calibration Mode and create new calibration settin...

Page 23: ...Return jack on the front panel of the HT 3000P 1A AC 4 In the calibration mode hit the Reset button until L1 is displayed on the front panel 5 Once L1 is displayed press the Test button once and 80 w...

Reviews: