16
Calibration and Software Version Information
This will allow the user to see the version of the software as well as who performed the last
calibration.
1.
Turn off the HT-3000P-1A-AC tester.
2.
Hold in the Reset button while turning on the tester.
3.
The meter will display 3 items:
A) The model number of the tester.
B) The version of the software.
C) Laboratory number to designate who performed the last calibration:
(1= Compliance West USA, 2= another company)
Voltage Calibration Adjustment
Use the following procedure to calibrate the output voltage.
1.
Ensure that all test leads are removed from the HT-3000P-1A-AC.
2.
Turn OFF the HT-3000P-1A-AC tester.
3.
To enter the Calibration Mode you MUST follow the following sequence:
- Push and hold in both the Test and Reset buttons.
- Turn on the HT-3000P-1A-AC tester.
- Release the Test button, release the Reset button, press and hold the Reset button, press
and hold the Test button, release the Test button, and then release the Reset button.
4.
If the correct sequence was entered, the display will read "Sure", if not, start again at step 2
above.
5.
While "Sure" is displayed on the screen you can:
A) Press Reset to exit out of the Calibration Mode and keep all of the currently
programmed calibration settings or
B) Press Test to enter the Calibration Mode and create new calibration settings. (Be
sure you want to enter the Calibration Mode as this will change the laboratory number so
it will show the calibration was not performed by Compliance West USA).
6.
Once the Calibration Mode has been entered, the Reset button toggles between the 5
calibration menus: Volt, V1, V2, curr, and bars. Make sure "V1" is displayed on the meter.
7.
Hook up a calibrated 1000:1 voltage probe and a calibrated DMM between output voltage
leads.
8.
Turn the Voltage Adjust on the rear panel to minimum (counterclockwise). Press the Test
button and 3000 will be displayed on the front panel meter. Be careful as the HT-3000P-
1A-AC will be putting out voltage at this point.
9.
Turn the Voltage Adjust on the rear panel clockwise until the output on the DMM equals
3000 volts AC and press the Test button. The display will now display "V1" again.
10. Press the Reset button until "V2" is displayed on the front panel meter.
11. Turn the Voltage Adjust on the rear panel to minimum (counterclockwise). Press the Test
button and 500 will be displayed on the front panel meter. Be careful as the HT-3000P-1A-
AC will be putting out voltage at this point.