10 

Initial Checkout Procedure 

 

The following procedure will verify that the HT-3000P-1A-AC is working correctly. We 

recommend that this procedure be conducted periodically to ensure proper operation of the tester.   
 

CAUTION 

High  voltage  (up  to  3000  Volts  AC)  generated  by  the  HT-3000P-1A-AC  is 
exposed during this test.  A risk of shock exists.  Exercise care when using the 
HT-3000P-1A-AC. 
 

Performance Test 

 

1. 

Connect the tester to a proper source of supply. Turn the HT-3000P-1A-AC on. 

2. 

Disconnect all test leads and set the variac at minimum voltage. 

3. 

Push  the  red  RESET  button.  Verify  the  yellow  TEST  indicator  is  lit  and  then  push  the 
TEST button. 

4. 

The  tester  should  conduct  a  test  sequence.  At  test  termination,  the  Full  Voltage  and  Red 
RESET LED’s should be lit, Push the RESET button. 

5. 

Connect  the  RETURN  lead  and  the  HIGH  VOLTAGE  TEST  lead  to  the  HT-3000P-1A-
AC. Set duration to more than 15 seconds. 

6. 

Press  and  hold  the  RESET  button,  until  the  leakage setting appears  on the display, verify 
the correct leakage current setting. 

7. 

Connect  the  RETURN  lead  to  the  HIGH  VOLTAGE  TEST  lead  (make  a  short  circuit), 
Keep clear of the wire while the test is in progress

8. 

Push the red RESET button and then push the yellow TEST button. 

9. 

Increase the current using the voltage variac knob until it reaches the leakage set point. 

10.  The  tester  should  conduct  a  test  sequence,  terminated with a buzzer.   At test termination, 

the Excess Leakage and red RESET LED’s should be lit. 

11.  Set duration to the desired test time. 

 

 

 

 
 

Passage  of  these  tests  indicates  that  the  HT-3000P-1A-AC  is  functioning  properly  and 

that it is safe to use. If the results of the performance test are not in accordance with the above, 
service  is  required.  Remove  the  HT-3000P-1A-AC  from  service  and  contact  Compliance  West 
USA, Inc. for servicing information. 
 

Setting up the HT-3000P-1A-AC  

 

This  section  describes  procedures  for  setting  the  Pass/Fail  leakage  current  level,  high 

voltage level, and high voltage test time. The HT-3000P-1A-AC is calibrated as shown below at 
the factory to be usable without adjustment in the majority of applications. If the factory settings 
are acceptable, you may skip this section.   
 

 

 
 
 
 

Summary of Contents for HT-3000P-1A-AC

Page 1: ...HT 3000P 1A AC Dielectric Withstand Tester 0 3000 Volts AC Output Instruction Manual...

Page 2: ......

Page 3: ......

Page 4: ...micro controlled circuit design and ergonomic front panel and represents the latest in high voltage production line testing To fully appreciate all the features of your new meter we suggest that you t...

Page 5: ...ii...

Page 6: ...al Checkout Procedure 10 Performance Test 10 Setting up the HT 3000P 1A AC 10 Factory Settings 11 Display of Leakage Limit and Duration settings 11 Adjustment of the Leakage Current Level 11 Adjustmen...

Page 7: ...1...

Page 8: ...for conducting the dielectric withstand test should be as remote as possible from normal production line activities Only personnel actually conducting the test should be allowed in the area and it sho...

Page 9: ...er s return lead is connected to ground potential and when properly connected to the equipment being tested it will guard against the operator contacting high voltage Always make sure the return lead...

Page 10: ...being tested are still failing the primary circuit capacitance of the equipment may be too high to allow the AC dielectric withstand test to be used This is due to the AC voltage charging all capacit...

Page 11: ...sing the RESET Button The test will not begin until the TEST Button is pushed If a failure is encountered the high voltage is immediately shut down a buzzer sounds and any voltage stored in the equipm...

Page 12: ...s Tamperproof operation available ENVIRONMENTAL Operating Temperature 15 40 C Relative Humidity Range 0 90 non condensing GENERAL Input power requirements 240V 60 Hz 15A single phase Weight 180 lbs ap...

Page 13: ...e during shipment We recommend that the instrument be surrounded by at least three inches of shock absorbing material on all sides of the container AC Line Voltage Requirements AC line voltage require...

Page 14: ...8 Figure 2 Controls Indicators Connectors Model HT 3000P 1A AC Front Panel...

Page 15: ...visual indication to the operator of the actual output voltage 9 Hipot Pass LED At the end of the test if no Excess leakage was detected the green hipot pass led will light and the test will terminate...

Page 16: ...y verify the correct leakage current setting 7 Connect the RETURN lead to the HIGH VOLTAGE TEST lead make a short circuit Keep clear of the wire while the test is in progress 8 Push the red RESET butt...

Page 17: ...to 1000 mA To set the leakage limit follow the steps below 1 Connect the HT 3000P 1A AC to a correctly rated source of supply and turn the power switch to the I or ON position Push the RESET button T...

Page 18: ...ng with testing the HT 3000P 1A AC tester should be set up for production line testing See Section above before continuing CAUTION High voltage is generated by the HT 3000P 1A AC Although the chassis...

Page 19: ...ce West USA is available Phone 800 748 6224 Hours 8 00 AM 4 00 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D San Diego C...

Page 20: ...the calibration procedure is a DMM able to read true rms 0 3000 Vac 1 and a 3kohm 3600 w 1 resistor Service Information The HT 3000P 1A AC is warranted to the original purchaser for a period of 1 year...

Page 21: ...Calibration adjustment calibrates the voltage output to agree with the meter reading 2 Leakage Current calibration adjustment calibrates the internal current meter to obtain the real current flowing b...

Page 22: ...you can A Press Reset to exit out of the Calibration Mode and keep all of the currently programmed calibration settings or B Press Test to enter the Calibration Mode and create new calibration settin...

Page 23: ...Return jack on the front panel of the HT 3000P 1A AC 4 In the calibration mode hit the Reset button until L1 is displayed on the front panel 5 Once L1 is displayed press the Test button once and 80 w...

Reviews: