Initial Checkout Procedure
The following procedure will verify that the HT-3000P-1A-AC is working correctly. We
recommend that this procedure be conducted periodically to ensure proper operation of the tester.
CAUTION
High voltage (up to 3000 Volts AC) generated by the HT-3000P-1A-AC is
exposed during this test. A risk of shock exists. Exercise care when using the
HT-3000P-1A-AC.
Performance Test
1.
Connect the tester to a proper source of supply. Turn the HT-3000P-1A-AC on.
2.
Disconnect all test leads and set the variac at minimum voltage.
3.
Push the red RESET button. Verify the yellow TEST indicator is lit and then push the
TEST button.
4.
The tester should conduct a test sequence. At test termination, the Full Voltage and Red
RESET LED’s should be lit, Push the RESET button.
5.
Connect the RETURN lead and the HIGH VOLTAGE TEST lead to the HT-3000P-1A-
AC. Set duration to more than 15 seconds.
6.
Press and hold the RESET button, until the leakage setting appears on the display, verify
the correct leakage current setting.
7.
Connect the RETURN lead to the HIGH VOLTAGE TEST lead (make a short circuit),
Keep clear of the wire while the test is in progress.
8.
Push the red RESET button and then push the yellow TEST button.
9.
Increase the current using the voltage variac knob until it reaches the leakage set point.
10. The tester should conduct a test sequence, terminated with a buzzer. At test termination,
the Excess Leakage and red RESET LED’s should be lit.
11. Set duration to the desired test time.
Passage of these tests indicates that the HT-3000P-1A-AC is functioning properly and
that it is safe to use. If the results of the performance test are not in accordance with the above,
service is required. Remove the HT-3000P-1A-AC from service and contact Compliance West
USA, Inc. for servicing information.
Setting up the HT-3000P-1A-AC
This section describes procedures for setting the Pass/Fail leakage current level, high
voltage level, and high voltage test time. The HT-3000P-1A-AC is calibrated as shown below at
the factory to be usable without adjustment in the majority of applications. If the factory settings
are acceptable, you may skip this section.
Summary of Contents for HT-3000P-1A-AC
Page 1: ...HT 3000P 1A AC Dielectric Withstand Tester 0 3000 Volts AC Output Instruction Manual...
Page 2: ......
Page 3: ......
Page 5: ...ii...
Page 7: ...1...
Page 14: ...8 Figure 2 Controls Indicators Connectors Model HT 3000P 1A AC Front Panel...