45
•
Mount the
automatic water inlet pump 46
on its base, strapping it to the supports.
•
Fit the branches (
Y
) and all connecting pipes.
•
Fit the external filling and waste pipes; the end of the inlet pipe is cut in a particular
manner to prevent a suction effect when it rests on the bottom of the tank.
The waste pipe is provided with a quick coupling male attachment.
•
Test tightness.
•
In the Technician Menu, Accessories option (option 16), set External Charge ON;
check operation of the sterilizer. Close and restart the unit.
Automatic inlet
Summary of Contents for Sterilclave 6 B Speedy
Page 2: ...2...
Page 6: ...6 Graphical reference representation Mod 6 B SPEEDY Contents...
Page 7: ...7 Contents...
Page 8: ...8 Contents...
Page 57: ...57 Diagrams and wirings...