Water in the chamber after the cycle
Proceed as follows if there is still water inside the
chamber 09
:
•
Check that a cycle with drying phase has been selected. Otherwise, the presence
of water or humidity in the chamber is a normal condition.
•
Check the cleanliness and correct positioning of the
drain filter 20
(see USE AND
MAINTENANCE MANUAL, chapter MAINTENANCE).
•
Check again the inclination of sterilizer (see USE AND MAINTENANCE MANUAL,
chapter INSTALLATION).
•
Check that the direct drain pipe is clear (it must NEVER be immersed in water).
Pressure rises too slowly
Proceed as follows if the pressure, rises too slowly at the end of the cycle:
•
Check the
air sterilization filter 15
and replace it, if needed. Note that this filter, as
many other components, has a specific life (see USE AND MAINTENANCE MANUAL,
chapter MAINTENANCE). NEVER run sterilization cycles without this filter or if the filter
is damaged or perforated. The load could be contaminated by ambient air.
Poor drying
Proceed as follows if drying is not satisfactory at the end of the cycle:
•
Check that a cycle with drying phase has been selected and that the selected
cycle is suitable for the material to be sterilized (see USE AND MAINTENANCE
MANUAL, chapter OPERATING USE).
•
Check that the total maximum load is not exceeded.
•
Check that load is prepared correctly.
•
Check the cleanness and correct positioning of the
drain filter 20
.
•
Check the
air sterilization filter 15.
•
Check again the inclination of sterilizer (see USE AND MAINTENANCE MANUAL,
chapter INSTALLATION).
•
Check that the valve
EV5
(see wiring diagrams) opens during the drying phase. This
valve is normally closed but allows air peaks in during the drying phase.
Steam leakage from the safety valve
A steam leakage from the
chamber safety valve 12
o
12A
can occur in various conditions:
•
A suitable safety valve must be fitted if the sterilizer is installed at more than
1500 meters above sea level (see USE AND MAINTENANCE MANUAL, chapter
TECHNICAL SPECIFICATIONS).
•
If the unit is installed at an altitude below 1500 meters above sea level and/or the
problem repeatedly occurs, check the valve efficiency or replace it.