Disconnect the connector corresponding to the temperature probe you want to calibrate
from the
electronic power board 34.
Disconnect connector CN12 for sensor PT1,
connector CN13 for sensor PT2 or connector CN14 for sensor PT3.
Connect the calibration tool 2-pin cable to the connector CN12, CN13 or CN14 of the
power cable 34.
Switch pole number 1 of the temperature selector to ON, which
corresponds to a 1000 Ohm sample coil (0°C).
Wait for at least 10 seconds and press ENTER.