10
Use the
↑↓
arrows to scroll the list and highlight the required option. Press ENTER to access
the selected option.
Press
Å
to go back to the MENU page.
Inputs
The status of the sterilizer inputs can be displayed:
•
S1 is the clean water tank
10
safety probe
•
S2 is the clean water tank
10
minimum level probe
•
S3 is the clean water tank
10
maximum level probe
•
S4 is not used in 6 B SPEEDY model
•
S5 is the
maximum level probe 43B
of the
steam generator 47
•
S6 is the
minimum level probe 43A
of the
steam generator 47
•
S7 is not used
•
PT1 is the
temperature probe 39
in the
chamber 09
•
PT2 is the
temperature probe 39A
in the
steam generator 47
•
PT3 is the
temperature probe 39B
on the external coil
•
Pressure Transducer 38
•
Conductivity meter
•
Micro-switch M1 36A
: door locked
•
Micro-switch M2 36B:
door pre-closure
Use the
↑↓
arrows to scroll the list and press
Å
to go back to the Technician Menu.