20
Accessories
Select Accessories (option 16) and press ENTER. The external charge value (ON/OFF) will
appear. Press
↑↓
to change the value, press ENTER to confirm and move to the next
option.
Press
Å
to go back to the Technician Menu without making changes to the accessory
settings.
Delete maintenance
Press YES to delete all maintenance interventions performed on the sterilizer from the
machine memory and reset the remaining day and cycle counters for all components.
Press NO to keep the maintenance log and go back to the Technician Menu.
Note: After installing the sterilizer, delete the maintenance log to reset the remaining days
and cycles for all components.
Technician Menu
Summary of Contents for Sterilclave 6 B Speedy
Page 2: ...2...
Page 6: ...6 Graphical reference representation Mod 6 B SPEEDY Contents...
Page 7: ...7 Contents...
Page 8: ...8 Contents...
Page 57: ...57 Diagrams and wirings...
Page 58: ...58 Wiring diagram 6 B SPEEDY Diagrams and wirings...