Installation of the SPEEDY WATER demineralization system
The SPEEDY WATER demineralization system should be used to connect the SterilClave 6/18/24
sterilizers to the mains water in automatic water inlet configuration B (see USE AND
MAINTENANCE MANUAL, chapter CONNECTIONS).
The SPEEDY WATER demineralization system consists of:
•
conductivity meter
to be fitted on the sterilizer;
•
purification block B
(ionic exchange resins), easy to substitute;
•
support bracket A
for the above mentioned block, to be attached to the wall.
Water quality check
The conductivity meter allows the sterilizer to constantly check that the quality of the
water coming from the purification block is suitable for sterilization.
In particular, when the first threshold (15 µS/cm, maximum conductivity recommended
under EN 13060 for feedwater) is exceeded, when in STAND BY mode the sterilizer displays
the message “Check water quality”, to remind that resins replacement is needed.
When the second threshold (30 µS/cm) is exceeded, the INSUFFICIENT WATER QUALITY
ALARM stops water coming in automatically from the purification block and warns that it is
necessary to change the resins.
Purification block
By treating mains water with a flow of about 650
µ
S/cm the
purification block B
is able to
produce 250 liters of water below the first threshold (15 µS/cm) and another 50 before
reaching the second (30 µS/cm) threshold, for a total of 300 liters (
Æ
600 cycles).
Technical data
Flow rate
between 0.5 and 0.7 l/min
Size H x W x D
520 x 160 x 90 mm
Operating weight
4.7 kg
Mains water requirements
pressure between 1 and 4.5 bar
Support bracket
The
support bracket A
is equipped with rapid connections for connecting to the mains
water (inlet) and to the sterilizer (outlet). These fittings are suitable for 8 mm external
diameter pipes. Four 7 mm holes for wall mounting are also available.
Bracket technical data
Size H x W x D
270 x 160 x 180 mm
Operating weight
0.9 kg
Maintenance of the SPEEDY WATER demineralization system
When the
purification block B
needs to be replaced, follow these steps:
•
Switch OFF the sterilizer, close the mains water tap and place a container below
the
support bracket A
to collect any water spillage.
•
Lift the front cover of the
support bracket A
.
•
Remove the
metal blocking plate D
.
•
Detach the
purification block B
from the support and install the new one as
described below.