48
•
Replace the
automatic water inlet pump 46
and the connection pipes with the
solenoid valve EV7
(mounted on its support and with the outlet hose holders and
the inlet L-joint) and the pipes supplied. Use the teflon pipe to connect the internal
joint on the back of the sterilizer with the
solenoid valve EV7
(inlet 1) and the silicone
pipe to connect the
solenoid valve EV7
(outlet 2) to the
clean water tank 10
.
•
Insert the
probe support E
in the silicon pipe and fasten the ends with the pipe
clamps provided. Connect to the pins using the appropriate screws and ready to
use, clamped blue cables with eyelets to the
filler 14
. Clamp the
probe support E
to
the supplied support screwed into the
slit J
, making sure there is no contact
between the pins and other parts of the sterilizer.
•
Open the mains water.
•
Test the sterilizer and the joints.
•
Check operations, close the mains water tap and install the sterilizer again.