PT1 / PT2 / PT3 probe calibration
This procedure is used to complete the calibration of the temperature probes PT1, PT2
and PT3.
This operation must be performed whenever a temperature probe is replaced.
Press +/- to enter the value 0°C for the temperature probe shown on the label or on the
wire sheath (e.g.: if a value of 1000.2 is shown on the PT1 sensor label, press + twice to
display 1000.2). Press ENTER to confirm.
Press
Å
to go back to the Technician Menu without changing the calibration.
PT1 / PT2 / PT3 correction
This procedure is used to correct the temperature value displayed during the sterilization
process. For example, if using an external sampling tool a temperature lower than the one
displayed is measured in the chamber during the sterilization process, the temperature
may be corrected as follows:
Press +/- to enter the correction which will change the temperature shown on the display.
Press ENTER to confirm.
Press
Å
to go back to the Technician Menu without changing the value.
Technician Menu