7
1.4
Electrical Requirements
THIS INSTRUMENT MUST BE GROUNDED
Before connection please ensure that the line supply corresponds to the power requirements below:
Power
Supply requirements
65 W
100 V - 230 V ~ 50/60 Hz
The equipment is provided with a power supply unit and three power cables consisting of a UK 3-pin and a “Schuko”
2-pin plug for 230 V installations and a NEMA 5-15 plug for 120 V installations.
Choose the power cable appropriate for your electrical installation and discard the others. Should one of the power
cables be suitable for connecting to the power supply, replace the plug with a suitable alternative..
THIS OPERATION SHOULD ONLY BE UNDERTAKEN BY A QUALIFIED ELECTRICIAN.
NOTE:
Refer to the equipment rating plate to ensure that the plug and fusing are suitable for the voltage and
wattage stated. The wires in the mains cable are as follows:
230 V a.c.
120 V a.c.
HOT/LIVE - BROWN
BLACK - HOT/LIVE
NEUTRAL - BLUE
WHITE - NEUTRAL
EARTH - GREEN/YELLOW
GREEN - EARTH
The appropriate power cable and power adaptor combination should be connected to the equipment BEFORE
connection to the mains supply. Should the mains lead require replacement, a cable of 1.25mm² (AWG16) of
harmonised code H05VV-F connected to an IEC320 plug should be used.