During scanning (blank, standard or sample)
The following errors can appear whenever the optical hardware is used. Each error message will be attached to
the activity which was just performed.
The error messages are displayed as a notification on screen with an option to display more information about the
error. If multiple errors have been detected, they will be displayed in the more information.
The ‘
blank / standard / sample
’ failed due to a potential hardware fault
A fault has been found whilst taking the reading. If an accessory is connected, check the cable is firmly
connected and turn the instrument on and off again before trying again. If this continues to occur contact
Cole-Parmer support and quote error message 112.
The ‘
blank / standard / sample
’ failed due to a potential hardware fault
A fault has been found whilst taking the reading. If an accessory is connected, check the cable is firmly
connected and turn the instrument on and off again before trying again. If this continues to occur contact
Cole-Parmer support and quote error message 113.
The ‘
blank / standard / sample
’ failed because the lamp has been disabled due to being too warm
The instrument has disabled the lamp because the unit is too warm. Check the air vents for any obstructions
which may be interrupting airflow, and turn the unit off to allow it to cool down. If this continues to occur
contact Cole-Parmer support and quote error message 114.
The ‘
blank / standard / sample
’ failed because a potential hardware fault detected with the fan
The fan has dropped below its normal operating speed. Check the air vents for any obstructions which may be
interrupting airflow, and then turn the instrument off and on again before trying again. If this continues to
occur contact Cole-Parmer support and quote error message 115.
The ‘
blank / standard / sample
’ failed due to potential hardware fault
Turn the instrument on and off again and try again. If this continues to occur, contact Cole-Parmer support
and quote error message 116.
The ‘
blank / standard / sample
’ failed because the lamp is too hot
Check the air vents for any obstructions which may be interrupting airflow and turn the unit off to allow it to
cool down. If this continues to occur, contact Cole-Parmer support and quote error message 117.
The following warning messages will be displayed next to results when detected. They are indicated by an icon,
which when pushed will display a dialog containing the warning information.
Lamp warm up warning
The lamp is still warming up. This can take up to 30 minutes after turning the instrument on, or after entering
lamp safe mode. Readings taken during this time may be inaccurate.
Calibration data lost warning
This reading may be inaccurate as a hardware problem has affected calibration data.