59
Section 15 -
Technical Specification
15.1
General Specification
7410
7415
W
avelength
Range
320 to 1000nm
198 to 1000nm
Resolution
1nm
Accuracy
± 2nm
Repeatability
± 0.5nm
Spectral bandwidth
5nm
Photometrics
Transmittance
0 to 199.9%
Absorbance
- 0.300 to 2.500 A
Accuracy
± 1%T, ± 0.01 Abs at 1.000 Absorbance
Resolution
0.1%T, 0.001 A
Stray light
<0.5% at 340nm
<0.5% at 340nm and 220nm
Stability
<0.002 Abs/hr after 30 minute warm up
<0.001 Abs/hr without warm up
Concentration
Range
-300 to 9999
Calibration
Blank with a single standard or factor
Factor
0.001 to 10000
Standard
0.001 to 1000
Quantitation
Range
-300 to 9999
Calibration
Blank with up to 20 standards
Curve fit algorithms
Linear through zero, linear, quadratic through zero and quadratic
Kinetics
Measurement Time
2 to 9999 seconds
Calibration
Blank with a single standard or factor
Display
Graphical
Analysis
Concentration, rate of change, initial and final absorbance/%T
Spectrum
Range
320 to 1000nm
198 to 1000nm
Scan interval
Selectable 1, 2, 5 or 10nm
Analysis
Absorbance or % transmittance and peaks and valleys and area under curve
Other
Beam height
15mm
Light source
Tungsten halogen lamp
Xenon lamp
Lamp save
Yes
Not applicable
Internal memory
10GB for results and methods
External memory
Limited by attached mass storage device
Removable media
USB
Outputs
USB Type A x2, USB Type B x1, Ethernet connection (RJ-45)
Power supply unit
24 V d.c. 2.5 A
Size (w x d x h)
280 x 500 x 156mm
Weight
9.0kg
Warranty
3 years (including xenon lamp)