Connect CC-Link Wires (Optional)
Note
: The CC-Link terminals are only available on the CIO-MICRO-CC I/O module.
1. Determine how CC-Link devices will be connected to the I/O module’s CC-Link terminals.
2. Use a screwdriver to loosen the appropriate screw terminals.
3. Insert the CC-Link wires (using a CC-Link specified cable) to the CC-Link terminals and the other end of the
cables to the applicable CC-Link devices. Refer to
CC-Link Terminal Block Assignments
for terminal
block pin assignments. Refer to
for common wiring configurations.
4. To reduce emissions, attach a Steward 28A0640-0A2 ferrite around the CC-Link wire bundle, as close to the
connector as possible.
Note
: The CC-Link network is daisy-chained and requires a terminal resistor for the first and last devices in
the chain. Make certain that your connections are correct. Refer to the CC-Link website for more information
and specification details.
5. Tighten the screw terminals with the screwdriver to secure the wire leads in the terminal block; the maximum
torque is 0.1921 Nm (1.7 in-lb).