Maximum Scan Rates for Measuring Fields
Maximum Acquiring Rate Depending on the Measuring Field Number
+0
+1
+2
+3
+4
+5
+6
+7
0
144Hz
192Hz
192Hz
192Hz
284Hz
284Hz
284Hz
552Hz
8
181Hz
239Hz
239Hz
239Hz
354Hz
354Hz
354Hz
680Hz
16
181Hz
239Hz
239Hz
239Hz
354Hz
354Hz
354Hz
680Hz
24
181Hz
239Hz
239Hz
239Hz
354Hz
354Hz
354Hz
680Hz
32
241Hz
318Hz
318Hz
318Hz
469Hz
469Hz
469Hz
892Hz
40
241Hz
318Hz
318Hz
318Hz
469Hz
469Hz
469Hz
892Hz
48
241Hz
318Hz
318Hz
318Hz
469Hz
469Hz
469Hz
892Hz
56
241Hz
318Hz
318Hz
318Hz
469Hz
469Hz
469Hz
892Hz
64
362Hz
476Hz
476Hz
476Hz
694Hz
694Hz
694Hz
1200Hz
72
362Hz
476Hz
476Hz
476Hz
694Hz
694Hz
694Hz
1200Hz
80
362Hz
476Hz
476Hz
476Hz
694Hz
694Hz
694Hz
1200Hz
88
362Hz
476Hz
476Hz
476Hz
694Hz
694Hz
694Hz
1200Hz
96
552Hz
552Hz
1030Hz
1030Hz
1030Hz
1030Hz
1030Hz
1030Hz
104
680Hz
680Hz
1200Hz
1200Hz
1200Hz
1200Hz
1200Hz
1200Hz
112
892Hz
892Hz
1200Hz
1200Hz
1200Hz
1200Hz
1200Hz
1200Hz
120
1200Hz
1200Hz
1200Hz
1200Hz
1200Hz
1200Hz
1200Hz
1200Hz
Recommended Shutter Times
The value of
exposure
(in milliseconds) should be set depending on the material that is being scanned. The
recommendations are as follows:
Recommended shutter times (approximate values)
Target material
Shutter Time
White paper/plastic
10 - 50μs
Colored plastic
50 - 100μs
Metallic surfaces
0.1 - 1ms
Black plastic/rubber
0.5 - 1ms
The measuring field can be restricted by omitting complete matrix areas in order to suppress interfering image ranges.
Measuring field
and
measuring range
must be clearly differentiated in practical use. The measuring field is related to
the matrix and the measuring range is related to the measuring object (the object space). Both do not have to match on
account of the optical mapping and the definitions. The measuring field “standard” is larger than the measuring range
“standard”.
The DS900 Series Sensors are distinguished by:
l
A laser line with 20° opening angle (measuring range 25 mm) respectively 25° opening angle (measuring ranges
50 mm and 100 mm).
l
The receiver has a smaller opening angle (view angle) than the laser line.
l
Centered measuring field (symmetrical to the center axis).
l
The high resolution sensor image matrix has 1280 x 1024 pixels. The measuring field geometry is fixed.
32
Measuring Field Selection
Summary of Contents for DS900 Series
Page 1: ...DS900 Series Displacement Sensors Reference 02 21 2018 Version 9 3 0 5...
Page 14: ...Dimensions DS910B 14 Dimensions DS910B...
Page 17: ...Dimensions DS925B 17 Dimensions DS925B...
Page 18: ...18 Dimensions DS925B...
Page 28: ...28 RS422 Synchronization...
Page 31: ...31 Measuring Field Selection...
Page 42: ...January 2018...