Your Hipot tester
89
‘Q’
1.30kV
‘R’
1.35kV
‘S’
1.40kV
‘T’
1.45kV
‘U’
1.50kV
‘V’
1.55kV
Note that not all combinations of ‘Base Target Voltage’ and ‘Target
Offset Voltage’ are Valid. AC Hipot Tests have a 5.00kV range, while
DC Hipot tests have a 0.1 - 6.00kV range and DC IR Tests have
discrete output voltages of 0.25kV, 0.5kV and 1.0 kV.
HAL's test code editor will not generate invalid combinations and all
invalid entry or scanned testcodes will be rejected.
A 0.00kV output voltage is intended to be used as a ‘rest’ period
between other tests in a sequence. Such sequences may be linked or
repeated (See digit 11).
Digit 5 – Ramp to Target Time
Digit 6 – Hold Time
Digit 7 – Ramp to Zero Time
Digits 5,6 and 7 represent times during which the out voltage is ramped
to the target voltage, held at the target voltage and finally ramped down
to zero. All the times are in seconds.
Digit 5
Ramp to Target
Time
Digit 6
Hold Time
Digit 7
Ramp to Zero
Time
‘0’
0.0
0.0
0.0
‘1’
0.1
0.1
0.1
‘2’
0.2
0.2
0.2
‘3’
0.3
0.3
0.3
‘4’
0.4
0.4
0.4
‘5’
0.5
0.5
0.5