Your Hipot tester
19
Default Sequence
When you first enter this menu all of the previous values are displayed
and one of the field descriptions is highlighted.
Use the encoder highlight a field to be changed and press the
Rotate
Field
fast key to activate the encoder. Rotate Field will be highlighted.
Turning the encoder will now cycle through all the available values for
the field. To accept desired value press Rotate Field again to deactivate
encoder for further field value change (highlight deselected). Again use
encoder (or press
Next
or
Previous Field
fast key) to move cursor to
navigate through to other fields including the three icons fields (see test
icon descriptions) for any other changes. Repeat process of
‘select/deselect’ the Rotate Field key & turning encoder to search for
values needed until all desired fields are changed to the user’s
requirements. Press
OK
and a message “storing system data” will be
displayed. And depending on the test conditions the test will begin with
the Meter screen displayed.
Description of Fields setting and Icons for Test Start conditions:
The first field (1
st
column from left) displays the
testcode
for the test.
The test type (eg. 50Hz Hipot) is shown directly below it. The tests
sequence may hold up to 5 unique tests. To add a second testcode for
example; highlight
SKIP
, press
Rotate Field
, turn encoder to select
values and press Rotate Field again to accept option. Follow through
the arrows to add/delete/edit any levels of testcodes sequence if