The unit goes back to the screen shown in
.
To move to the next parameter on the menu, press the
button.
Default value:
do not perform the test.
4�6�8�8� AVeRAge VOLTAge ReSeT
Screen used to reset the average voltage calculation.
The screen flashes between the symbol and the configured value. (
).
figure 56:Average voltage reset screens�
To edit the parameter, press the
button.
The options are:
Reset performed
Reset not performed
To move from one option to another, press the
button repeatedly.
When you see the option you want, validate by pressing the
button.
The unit goes back to the screen shown in
.
To move to the next parameter on the menu, press the
button.
Default value:
do not perform the test.
4�6�8�9� DISpLAy TeST
The test screen consists of all the segments of the screen lighting up in a sequence, allowing
you to verify the correct operation of all the indicator lights.
The screen flashes between the symbol and the configured value. (
)
figure 57:Display test screen�
To edit the parameter, press the
button.
The options are:
46
DHB-302 DHB-324
Instruction Manual