Charge/Discharge Controller 69200-1 User’s Manual
3-2
SCPI command settings:
1. CHANnel:SOURce 1
2. OUTPut:STATe OFF
3. SOURce:MODe CVC
--------------------------------
4. SOURce:CURR 1.000
--------------------------------
5. SOURce:VOLTage:CUTOFF 3.850
----------------
6. SOURce:TIME:CUTOFF 0
--------------------------
can be replaced with SOUR:ALL
7. SOURce:CURRent:CUTOFF 0.100
----------------
8. SOURce:VOLT 3.800
---------------------------------
9. SOURce:POW 4
---------------------------------------
10. SOURce:SLEW 1.00
----------------------------------
11. OUTPut:STATe ON
12. MEAS:OPER?
---------------------------------------
13. MEAS:STATe?
---------------------------------------
14. MEAS:CURR?
---------------------------------------
15. MEAS:VOLT?
---------------------------------------
16. MEAS:POW?
----------------------------------------
can be replaced with MEAS:ALL
17. MEAS:TIME?
---------------------------------------
18. MEAS:TEMP?
---------------------------------------
19. MEAS:AH?
------------------------------------------
20. MEAS:KWH?
---------------------------------------
3.3.8.7 CP Charge Mode Operation
Description: When the operating channel is 1 and the current slew rate is set to 1.00A/ms,
4W is fixed for charging until the voltage is 3.800V.
Single unit settings:
P
4.00
(Power limit must be setup)
V C U T
3.800
(Cutoff Voltage must be setup)
T I M E
0 (Testing time must be setup, “0” means no time limit)
S.R.
1.00
(Current Slew Rate must be setup)
V
3.850
(must > VCUT)
I
1.000
(Max. Current limit must be setup)
I C U T
0
(not necessary, set as “0”)