Programmable DC Electronic Load 63600 Series Operation & Programming Manual
5-90
5.3.2.20 VOLTAGE Subsystem
:STATic
:L2
:L1
SP
<NRf+>
?
SP
MAX
MIN
<NR2>
Suffix
<NR2>
<NR2>
:VOLTage
:ILIMit
:TYPE
:RESponse
SP
<NRf+>
?
SP
MAX
MIN
<NR2>
Suffix
<NR2>
<NR2>
SP
CURR/PSU / 0
VOLT/PSU / 1
?
SP
FAST / 0
NORMAL / 1
?
SLOW / 2
<CRD>
<CRD>
VOLTage:STAT:ILIMit
Type:
Channel-Specific
Description:
Set the current limit for constant voltage mode.
Setting Syntax:
VOLTage:STATic:ILIMit<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example:
VOLT:STAT:ILIM 3
Set the current limit to 3A in
constant voltage mode.
VOLT:STAT:ILIM MAX
Set the current limit to the
maximum value in constant
voltage mode.
VOLT:STAT:ILIM MIN
Set the current limit to the
minimum value in constant
voltage mode.
Query Syntax:
VOLTage:STATic:ILIMit?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Ampere]
Query Example:
VOLT:STAT:ILIM?
VOLT:STAT:ILIM? MAX
VOLT:STAT:ILIM? MIN
VOLTage:STATic:L1
Type:
Channel-Specific
Description:
Set the static load voltage in constant voltage mode.
Setting Syntax:
VOLTage:STATic:L1<space><NRf+>[suffix]