5-22
that enabled by Standard Event Status register are
logically ORed
to cause the Status Byte register ESB (bit 5) to be set. See
descriptions of these three registers in
Chapter
Setting Syntax:
*ESE<space><NR1>
Setting Parameters: <NR1>, 0 ~ 255
Setting Example: *ESE 48
This command enables the CME and EXE events for
the Standard Event Status register.
Query Syntax:
*ESE?
Return Parameters: <NR1>
Query Example:
*ESE?
This query returns the current setting for "Standard
Event Status Enable".
*ESR? Standard Event Status Register Query
Type:
Device Status
Description:
This query reads the Standard Event Status register. Reading the
register clears it. See detailed explanation of this register in
Chapter
Standard Event Status Event Register
Bit Position
7
6
5
4
3
2
1 0
Condition
PON 0
CME EXE DDE QYE 0 0
Bit Weight
128 64
32
16
8
4
2 1
Query Syntax:
*ESR?
Return Parameters: <NR1>
Query Example:
*ESR? Return the Standard Event Status register readings.
Return Example: 48
*IDN? Identification Query
Type:
System Interface
Description:
This query requests the Electronic Frame (63600) to identify itself.
Query Syntax
*IDN?
Return Parameters:<aard>
Query Example:
*IDN?
String
Information
CHROMA
Manufacture
63600-5
Model
636000000001
Serial number
1.00
Revision level of the primary interference
firmware
Return Example:
CHROMA,63600-5,636000000001,1.00
*OPC Operation Complete Command
Type:
Device Status
Description:
This command causes the interface to set the OPC bit (bit 0) of the
Standard Event Status register when the Electronic Frame (63600)
has completed all pending operations.
Setting Syntax:
*OPC
Setting Parameters: nil
*OPC? Operation Complete Query
Type: