Programmable DC Electronic Load 63600 Series Operation & Programming Manual
4-52
d. 5.0V: It sets the Dout High level to 5.0V.
8. DGNDC: Digital signal reference ground.
9. EXT_WAVE [1:10]: External wave input signals from the first channel
to the tenth channel, the input range is from 0 to 10V.
10. VMON [1:10]: Voltage monitor output signals from the first channel to
the tenth channel, the output range is from 0 to 10V.
11. IMON [1:10]: Current monitor output signals from the first channel to
the tenth channel, the output range is from 0 to 10V.
12. AGNDC: Analog signal reference ground.
The System I/O port on the 63600-2 Mainframe rear panel is a 25-pin connector (D-SUB
25pin male connector). It includes Analog signals: voltage and current monitor and external
wave input, and Digital System I/O signals. The Digital System I/O signals are TTL
Compatible. They are defined as follows.
Figure 4-11 63600-2 System I/O Port Connector
Table 4-9 63600-2 Pin Assignments of the System I/O Port Connector
Pin No.
Signal
Pin No.
Signal
1
EXT_WAVE1
14
EXT_WAVE2
2
EXT_WAVE3
15
EXT_WAVE4
3
AGNDC
16
VMON1
4
VMON2
17
VMON3
5
VMON4
18
IMON1
6
IMON2
19
IMON3
7
IMON4
20
SHORT1
8
SHORT2
21
SHORT3
9
SHORT4
22
DGNDC
10
TRIG_DIGI
23
TRIG_SEQ
11
LOAD_ON
24
DO1
12
DO2
25
DI1
13
DI2
1.
TTL High Level Voltage is 5V
。
2.
SHORT [1:4]: Short ON output signals from the first channel to the
fourth channel, TTL Level, Active High.
3.
TRIG_ SEQ: External trigger input signal to get to the next
sequence automatically. TTL Level, falling edge, pulse width≥1μs.
4.
TRIG_DIGI: External trigger input signal to be the trigger Source of
Digitizing Function. TTL Level, falling edge, pulse width ≥1μs.
5.
LOAD_ON: Load ON output signal, TTL Level, Active High.
6.
DI [1:2]: 2 bits of digital input signals, TTL Compatible.
DI1 and DI2 have External Load ON/OFF function. The user can
use this input signal to control the Load ON/OFF externally. If DI1
Summary of Contents for 63600 Series