Remote Operation
5-49
Setting Example: CURR:DYN:L1 20
Set the dynamic load parameter
L1 = 20A.
CURR:DYN:L1 10A
Set the dynamic load parameter
L1 = 10A.
CURR:DYN:L1 MAX
Set the dynamic load parameter
L1 = maximum value.
CURR:DYN:L1 MIN
Set the dynamic load parameter
L1 = minimum value.
Query Syntax:
CURRent:DYNamic:L1?[<space><MAX | MIN>]
Return Parameters:<NR2>,[Unit = Ampere]
Query Example:
CURR:DYN:L1?
CURR:DYN:L1? MAX
CURR:DYN:L1? MIN
CURRent:DYNamic:L2
Type:
Channel-Specific
Description:
Set the load current during T2 period for constant current dynamic
mode.
Setting Syntax:
CURRent:DYNamic:L2<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: CURR:DYN:L2 20
Set the dynamic load parameter
L2 = 20A.
CURR:DYN:L2 10A
Set the dynamic load parameter
L2 = 10A.
CURR:DYN:L2 MAX
Set the dynamic load parameter
L2 = maximum value.
CURR:DYN:L2 MIN
Set the dynamic load parameter
L2 = minimum value.
Query Syntax:
CURRent:DYNamic:L2?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Ampere]
Query Example:
CURR:DYN:L2?
CURR:DYN:L2? MAX
CURR:DYN:L2? MIN
CURRent:DYNamic:REPeat
Type:
Channel-Specific
Description:
Set the repeat count for constant current dynamic mode.
Setting Syntax:
CURRent:DYNamic:REPeat<space><NRf+>
Setting Parameters:<NRf+>, 0 ~ 65535, Resolution = 1, Unit = None
Setting Example:
CURR:DYN:REP 500
Set repeat count = 500
CURR:DYN:REP MAX
Set repeat count = maximum value.
CURR:DYN:REP MIN Set repeat count = minimum value.
Query Syntax:
CURRent:DYNamic:REPeat?[<space><MAX | MIN>]
Return Parameters:<NR1>, [Unit = None]
Query Example:
CURR:DYN:REP?
CURR:DYN:REP? MAX
CURR:DYN:REP? MIN
CURRent:DYNamic:RISE
Type:
Channel-Specific
Description:
Set the rising slew rate of current for constant current dynamic
mode.
Setting Syntax:
CURRent:DYNamic:RISE<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.