Programmable DC Power Supply 62000P Series Operating & Programming Manual
[ C U R R E N T S E T T I N G C A L I B R A T I O N]
S H O R T O U T P U T T E R M I N A L A N D
P R E S S [ E N T E R ]
S E T T I N G O U T P U T C U R R E N T 2 0 . 0 A
A C T U A L O U T P U T C U R R E N T = 1 9 . 9 9 2 2 A
S E T T I N G O U T P U T C U R R E N T 5 0 . 0 A
A C T U A L O U T P U T C U R R E N T = 4 9 . 9 8 1 2 A
]
3
[
S T A R T H I G H C U R R E N T R A N G E ( 2 0 A )
C A L I .
!
C O N N E C T T H E C U R R E N T
S H U N T
T O P R O P E R
R A N G E P R E S S [ E N T E R ]
W H E N R E A D Y
Figure 3-91
(8) Press ENTER will perform 20A calibration. The system will set the output current to
20.00A automatically and the cursor will stop at position [3] as Figure 3-92 shows. Input
the current (4 digits after decimal point) read by Current Shunt (DVM) and press
“
ENTER
” to confirm.
[ C
U
R R E N T S E T T I N G C A L I B
R
A
T I O
N
]
S H O R T O U T P U T T E R M I N A L A N D
P R E S S [ E N T E R ]
S E T T I N G O U T P U T C U R R E N T 2 0 . 0 A
A C T U A L O U T P U T C U R R E N T = 1 9 . 9 9 2 2 A
S E T T I N G O U T P U T C U R R E N T 5 0 . 0 A
A C T U A L O U T P U T C U R R E N T = 4 9 . 9 8 1 2 A
]
3
[
Figure 3-92
(9) Now the DC Power Supply will be set to off again and a message will be popped up to
remind user to reconnect to proper current shunt range. Set current shunt whose rating is
closest to but still cover 50A. For Prodigit 7530, use 250A shunt directly.
[ C U R R E N T S E T T I N G C A L I B R A T I O N]
S H O R T O U T P U T T E R M I N A L A N D
P R E S S [ E N T E R ]
S E T T I N G O U T P U T C U R R E N T 2 0 . 0 A
A C T U A L O U T P U T C U R R E N T = 1 9 . 9 9 2 2 A
S E T T I N G O U T P U T C U R R E N T 5 0 . 0 A
A C T U A L O U T P U T C U R R E N T = 4 9 . 9 8 1 2 A
]
3
[
S T A R T H I G H C U R R E N T R A N G E ( 5 0 A ) C A L I .
!
C O N N E C T T H E C U R R E N T
S H U N T
T O P R O P E R
R A N G E P R E S S [ E N T E R ]
W H E N R E A D Y
Figure 3-93
3-56