Programmable DC Power Supply 62000P Series Operating & Programming Manual
When set to FACTORY DEFAULT = NO the instrument will retain the last
configuration saved by user. On the contrary, if it is set to FACTORY DEFAULT =
YES all configuration settings will return to the factory default.
In the mean time, the screen will display the messages of DEVICE MODEL, SERIAL
NO. and FIRMWARE VERSION.
DEVICE MODEL
: Display the model no. as
62012P-80-60
in the above figure.
SERIAL NO.
: Display the serial no. as
000001
in the above figure.
FIRMWARE VERSION: Display firmware version as
00.01
and released date as
2005/01/01 in the above figure.
3. Press
“
ENTER
” to confirm.
4. Press
“
EXIT
” to return to MAIN PAGE.
3.3.7 CALIBRATION
Chroma 62000P Series DC Power Supplies have 4 calibration functions: (1) VOLTAGE: the
actual voltage output (CV mode) and its measurement accuracy, (2) CURRENT: the actual
current out (CC mode) and its measurement accuracy, (3) APG VOLTAGE: the actual
voltage output and its accuracy of analog V Monitor under analog voltage control mode (4)
APG CURRENT: the actual current output and its accuracy of analog I Monitor under analog
current control mode.
Follow the procedure below to enter into calibration mode:
1. In CONFIG Setup page, press “
6
”.
2. Press
“
ENTER
” to enter into CALIBRATION option as Figure 3-73 shows.
[ C A L I B R A T I O N ]
P A S S W O R D
:
Figure 3-73
3-46