Programmable DC Power Supply 62000P Series Operating & Programming Manual
[ A P G C U R R E N T C A L I B R A T I O N ]
C H E C K A P G C O N N E C T I ON A N D P R E S S [ E N T E R ]
( S E T ) I N P U T V O L T A G E F O R S E T T I N G = 0 . 5 V
A C T U A L A P G
I N P U T V O L T A G E
= 0 . 6 6 V
( S E T ) I N P U T V O L T A G E F O R S E T T I N G = 8 . 0 V
A C T U A L A P G
I N P U T V O L T A G E
= 8 . 6 8 V
]
1
[
]
2
[
Figure 3-102
(5) Press “
” again will ask users to input about 8.0V voltage signal (Pin 4), and press
“
ENTER
” the cursor will stop at position [2] as Figure 3-102 shows. Adjust the Power
Supply to 8V±0.2V and use DVM1 to measure the reading of Power Supply. Input the
voltage (2 digits after decimal point) read by DVM 1 to position [2] and press “
ENTER
”
to confirm.
(6) Press “
” again the system will set the output voltage of Pin 6 on the rear panel to
0.5V. The cursor stops at position [3] as Figure 3-103 shows. Input the voltage (2 digits
after decimal point) read by DVM 2 to position [3] and press “
ENTER
” to confirm.
[ A P G C U R RE N T C A L I B R A T I O N ]
C H E C K A P G C O N N E C T I ON A N D P R E S S
[ E N T E R ]
( M E A S . ) O U P U T V O L T AG E F OR ME A S U R E = 0 . 5 V
A C T U A L A P G OU P U T V O L T A G E
= 0 . 6 6 V
( M E A S . ) O U P U T V O L T AG E F OR ME A S U R E = 8 . 0 V
A C T U A L A P G OU P U T V O L T A G E
= 7 . 6 6 V
]
3
[
]
4
[
Figure 3-103
(7) Press “
” again the system will set the output voltage of Pin 6 on the rear panel to
8.0V. The cursor stops at position [4] as Figure 3-103 shows. Input the voltage (2 digits
after decimal point) read by DVM 2 to position [4] and press “
ENTER
” to confirm.
(8) The APG Voltage calibration is done once the above actions are completed. Press “
SAV
E
”
and “
ENTER
”to save the calibrated data as Figure 3-95 shows, or press “
EXIT
” to return
to Calibration page without saving.
(9) Press “
EXIT
” to return to MAIN PAGE.
3-62